Motor doesn't move 

  

 

1/ if shaft spins freely: check connections to motor and check voltage to 
motor 
2/ if shaft rubs or is frozen, relieve binding and retest 

Motor makes noise 

 

 

1) Make sure that the fan or blower wheel is not contacting its housing. 
Adjust the motor mounting bracket position to re-center the fan or blower 
wheel, if necessary. 
2) Check the fan or blower wheel for damage or out of balance condition. 
Replace the fan or blower wheel if it is damaged or out of balance. 
3) Turn the motor shaft to make sure that it spins freely. If it binds or the 
bearings make a rubbing or scrapping sound then replace the motor. 

 

OTHER 

Controller on at all times - "locked-up" 

 

 

1/ Adjust set point to room temperature.  If the light goes out but is still 
heating, replace the solid state relay.  
2/ turn unit off and on to reset 
3/ if cannot change any condition on the front panel, call Customer 
Service 

Controller timer resets on its own 

 

 

1/ confirm that power from wall is consistent and within specs    
2/ call Customer Service with serial number  

Front panel displays are off 

 

 

1/ Check for wire damage. 

Unit or wall fuse/circuit breaker is blown 

 

 

1/ check wall power source 
 2/ compare current draw and compare to specs on data plate  
3/ see what other loads are on the wall circuit 

Unit will not turn on 

 

 

1/ check wall power source 
2/ check fuse/circuit breaker on unit or in wall 
3/ see if unit is on, e.g., fan or heater, and just controller is off  
4/ check all wiring connections, esp. around the on/off switch   

Unit is smoking 

–out of box 

 

 

1/ put unit under vent and run at full power for one hour. 
2/ follow special burn in procedures in section 6.0. 

Contamination in chamber 

 

 

1/ see cleaning procedure in operator’s manual 
2/ develop and follow SOP for specific application; include definition of 
cleaning technique and maintenance schedule  

Contamination in sample 

 

 

1/ see “Contamination in chamber. 

SERVICE 

If  this  product  should  require  service,  contact  your  service  representative.    For  information  on 
where to reach customer service please see the front cover of this manual. Should return of this 
product  be  necessary  a  return  Authorization  number  must  be  obtained  and  the  product  shipped 
according  to  your  representatives  instructions  to  the  indicated  service  center.    If  a  return 
authorization  number  is  not  obtained  then  the  product  will  be  returned  back  to  the  sender.    To 
insure  prompt handling, the  return  authorization  number  should  be  placed  on  the  outside  of  the 
package  and  a  detailed  explanation  of  the  return  enclosed  with  the  item.    After  receiving  the 
Return  Authorization  number,  the  item  must  be  received  within  30  days  if  not  the  return 
authorization will be void.

 

Содержание SMO5CR-2

Страница 1: ...Previously Designated as CR1 2 INSTALLATION AND OPERATIONAL MANUAL 01 2013 4861580 P O Box 627 300 N 26th Avenue Cornelius Oregon 97113 EMAIL tech Shellab com INTERNET http www Shellab com 1 800 322...

Страница 2: ...NCE SECTION 9 0 TROUBLESHOOTING AND SERVICE SECTION 10 0 PARTS LIST UNIT SPECIFICATIONS SCHEMATICS This unit is a special purpose oven for professional industrial or educational use where the preparat...

Страница 3: ...and enter your claim on the form supplied by the carrier 1 2 Inspect for concealed loss or damage on the unit itself both interior and exterior If necessary the carrier will arrange for official inspe...

Страница 4: ...you should consult your manual for further description or discussion of a control or user item 2 2 Indicates Adjustable Temperature 2 3 Indicates AC Power ON 2 4 Indicates High Limit Safety 2 5 Indic...

Страница 5: ...c Avoid direct sun fast moving air currents heating cooling ducts and high traffic areas Allow a minimum of 30 cm between the unit and walls or partitions which might obstruct free air flow 3 3 Liftin...

Страница 6: ...own set point preventing sample loss or unit degradation Please note that it is not recommended that the unit be allowed to operate using only the High Limit to control temperature as temperature unif...

Страница 7: ...ealt with according to local regulations Your dealer can provide you with a power exhaust which greatly helps under these applications 5 3 Do not operate near noxious fumes 5 4 Do not place sealed or...

Страница 8: ...mbled and cleaned Make sure unit is unplugged and cooled down before disassembly When disassembling use an 8 32 nut driver for removing all the cap nuts Remove the bottom liner insert before removing...

Страница 9: ...9 Figure 2...

Страница 10: ...Once the down arrow has been pushed use the up and down arrow to achieve desired set point 3 After reaching desired set point the display should blink once to enter the value 4 To Auto Tune push the u...

Страница 11: ...C in 30 minutes and control for 1 hour at 100 C Then you want to ramp up to 150 in 30 Minutes and control for 2 hours Then you want to cool down to start Set Point with Power Exhaust Outlet activated...

Страница 12: ...and the display should read SENS This enters you in the Input Configuration Menu To navigate through the menu press the up and down arrows Press the up or down arrow repeatedly until you reach OFFS in...

Страница 13: ...en cleaning and assure all volatile or flammable cleaners are evaporated and dry before reattaching the unit to the power supply 8 2 Storage If the oven is to be turned off it can be reactivated witho...

Страница 14: ...irm with front panel lights that controller is calling for heat 7 high limit failure confirm with front panel lights that High Limit is operating correctly 8 wiring problem check all functions and com...

Страница 15: ...problem solved then follow instructions in Section 7 5 for correct setting 8 electrical noise remove nearby sources of RFI including motors arcing relays or radio transmitters 9 bad connection on tem...

Страница 16: ...other loads are on the wall circuit Unit will not turn on 1 check wall power source 2 check fuse circuit breaker on unit or in wall 3 see if unit is on e g fan or heater and just controller is off 4...

Страница 17: ...inlet 2800503 Heating Element 9570831 Hepa filter 9990503 High Limit Controller 1750571 High Limit Indicator Pilot Lamp 4650553 Leveling Feet 2700512 Main Temperature Controller 1750803 ON OFF Switch...

Страница 18: ...5 lbs Dimensions Exterior WxDxH in Interior WxDxH in SMO5CR 2 CR1 2 35 x 30 25 x 38 17 x 20 x 20 125 Capacity Cubic Feet SMO5CR 2 CR1 2 4 0 Temperature Range Uniformity Control SMO5CR 2 CR1 2 Amb 15 t...

Страница 19: ...19 WIRE DIAGRAM...

Страница 20: ...includes components not manufactured by Manufacturer is not covered by this warranty Any alteration or removal of the serial number on Manufacturer s parts will void this warranty Under no circumstanc...

Отзывы: