Shel lab SMO5CR-2 Скачать руководство пользователя страница 11

11

 

2. 

After you reach “CONF”, push the PGM Button again and use 

 Up 

and  Down  Arrows  to  navigate  until  you  reach  “PRO”  in  the  bottom 
display and push the “PGM” button again. 

3. 

The display should now read “FNCT” in the bottom display and “2” in 
the  top  display.    You  will  need  to  change  the  “2”  to  “0”.    To  do  that, 
push  the  “PGM”  button  once  and  the  “FNCT”  should  start  flashing.  
While  it  is  flashing,  use  the 

 

Down  Arrow  to  change  the  “2” to  “0”.  

When  the  “FNCT”  stops  flashing,  push  the  exit  button  repeatedly  to 
return  to  normal  display  (Actual  Value  on  Top  and  Setpoint  Value  on 
Bottom.) 

4. 

Now the control is setup as a Single Setpoint Control.  Just use the 

 

Down Arrows to change Setpoint Value. 

 

EXAMPLE PROGRAM 

Let’s say you want  to have the oven to ramp up to 100°C in 30 minutes 
and control for 1 hour at 100°C.  Then you want to ramp up to 150° in 30 
Minutes and control for 2 hours.  Then you want to cool down to start Set 
Point with Power Exhaust Outlet activated in one hour.  See Graph: 
 

EXAMPLE PROGRAM CHART

20

30

40

50

60

70

80

90

100

110

120

130

140

150

160

30 MINUTES INTERVALS

T

EM

PE

R

A

T

U

R

E

25

100

100

150

150

150

150

25

 

 
You would program the control like this: 
STEP1 

SP01=25° 

TPO01=00:30hr 

OC01-0000 

STEP2 

SP02=100° 

TPO02=01:00hr 

OC02-0000 

STEP3 

SP03=100° 

TPO03=00:30hr 

OC03-0000 

STEP4 

SP04=150° 

TPO04=02:00hr 

OC04-0000 

STEP5 

SP05=150° 

TPO05=01:00hr 

OC05-0100 

STEP6 

SP06=25° 

TPO06=00:00hr 

OC06-0000 

The OC parameter 0100 is the value that activates the Exhaust Outlet on 
the back of the oven. 

To  activate  the  program  that  is  entered  first  return  to  normal  display  by 

pushing  the 

Exit

  Button  repeatedly. 

    After 

Normal  Display

  is 

reached, push the up arrow 

 once.  The lower display should flash 

STrT

 

Содержание SMO5CR-2

Страница 1: ...Previously Designated as CR1 2 INSTALLATION AND OPERATIONAL MANUAL 01 2013 4861580 P O Box 627 300 N 26th Avenue Cornelius Oregon 97113 EMAIL tech Shellab com INTERNET http www Shellab com 1 800 322...

Страница 2: ...NCE SECTION 9 0 TROUBLESHOOTING AND SERVICE SECTION 10 0 PARTS LIST UNIT SPECIFICATIONS SCHEMATICS This unit is a special purpose oven for professional industrial or educational use where the preparat...

Страница 3: ...and enter your claim on the form supplied by the carrier 1 2 Inspect for concealed loss or damage on the unit itself both interior and exterior If necessary the carrier will arrange for official inspe...

Страница 4: ...you should consult your manual for further description or discussion of a control or user item 2 2 Indicates Adjustable Temperature 2 3 Indicates AC Power ON 2 4 Indicates High Limit Safety 2 5 Indic...

Страница 5: ...c Avoid direct sun fast moving air currents heating cooling ducts and high traffic areas Allow a minimum of 30 cm between the unit and walls or partitions which might obstruct free air flow 3 3 Liftin...

Страница 6: ...own set point preventing sample loss or unit degradation Please note that it is not recommended that the unit be allowed to operate using only the High Limit to control temperature as temperature unif...

Страница 7: ...ealt with according to local regulations Your dealer can provide you with a power exhaust which greatly helps under these applications 5 3 Do not operate near noxious fumes 5 4 Do not place sealed or...

Страница 8: ...mbled and cleaned Make sure unit is unplugged and cooled down before disassembly When disassembling use an 8 32 nut driver for removing all the cap nuts Remove the bottom liner insert before removing...

Страница 9: ...9 Figure 2...

Страница 10: ...Once the down arrow has been pushed use the up and down arrow to achieve desired set point 3 After reaching desired set point the display should blink once to enter the value 4 To Auto Tune push the u...

Страница 11: ...C in 30 minutes and control for 1 hour at 100 C Then you want to ramp up to 150 in 30 Minutes and control for 2 hours Then you want to cool down to start Set Point with Power Exhaust Outlet activated...

Страница 12: ...and the display should read SENS This enters you in the Input Configuration Menu To navigate through the menu press the up and down arrows Press the up or down arrow repeatedly until you reach OFFS in...

Страница 13: ...en cleaning and assure all volatile or flammable cleaners are evaporated and dry before reattaching the unit to the power supply 8 2 Storage If the oven is to be turned off it can be reactivated witho...

Страница 14: ...irm with front panel lights that controller is calling for heat 7 high limit failure confirm with front panel lights that High Limit is operating correctly 8 wiring problem check all functions and com...

Страница 15: ...problem solved then follow instructions in Section 7 5 for correct setting 8 electrical noise remove nearby sources of RFI including motors arcing relays or radio transmitters 9 bad connection on tem...

Страница 16: ...other loads are on the wall circuit Unit will not turn on 1 check wall power source 2 check fuse circuit breaker on unit or in wall 3 see if unit is on e g fan or heater and just controller is off 4...

Страница 17: ...inlet 2800503 Heating Element 9570831 Hepa filter 9990503 High Limit Controller 1750571 High Limit Indicator Pilot Lamp 4650553 Leveling Feet 2700512 Main Temperature Controller 1750803 ON OFF Switch...

Страница 18: ...5 lbs Dimensions Exterior WxDxH in Interior WxDxH in SMO5CR 2 CR1 2 35 x 30 25 x 38 17 x 20 x 20 125 Capacity Cubic Feet SMO5CR 2 CR1 2 4 0 Temperature Range Uniformity Control SMO5CR 2 CR1 2 Amb 15 t...

Страница 19: ...19 WIRE DIAGRAM...

Страница 20: ...includes components not manufactured by Manufacturer is not covered by this warranty Any alteration or removal of the serial number on Manufacturer s parts will void this warranty Under no circumstanc...

Отзывы: