APPENDIX TD
VACUUM LEAK DETECTOR
VL(R) ..
05/02/2018
- TD-1 -
Technical data
1. Electrical data
Electrical supply (without external signal)
230 V, 50 Hz, 50 W
24 VDC, 50 W
Switch contact load, terminal strips AS (5 and 6)
230 V
max: 230 V, 50 Hz, 200 VA
min:
20 mA
24 VDC
max: 24 VDC, 1 A
Switch contact load, dry relay contacts
max: 230 V, 50 Hz, 3 A
(terminals 11, 12 and 13)
min:
6 V / 10 mA
External fuse for the leak detector
max. 10 A
Overvoltage category
2
Protection type of housing,
Plastic
IP 30
Implementation VL …/p (sheet steel)
IP 54
2. Pneumatic data (requirements concerning the test measuring gauge)
Nominal size
min. 100
Accuracy class
min. 1.6
Scale end value
-600 mbar / -1000 mbar