- 21 -
SI 560A DIN SIZE WEIGHING INDICATOR
5-3-2. F-Function List
F-list
Subject
Default
Contents
101
Equipment No. setting – ID No.setting
01
01~99
102
Weight–Back up Mode
01
00:Normal mode
01: Weight Back up Mode(Zero)
02: Weight Back up Mode(Zero&Tare)
103
Weighing Data Save Method
(Refer page 26)
03
00: Manual: Whenever “Print” key input
01: Auto: At every steady states
02: Auto: At the first steady states
04: Manual&Auto: At every steady states
05:Manual&Auto: At the first steady
states
06: Manual&Auto: After weighing is
finished
104
Display Up-Date Speed
09
01:Slow(1time per 1sec)
~ 09:Fast(60times per 1sec)
106
No. 1 sub-display setting
01
00:P/N 01:SP1 02:SP2 03:SP3 04:SP4
05:Pass Count 06:Over Count 07:Under
Count 08:Error Count
107
No. 2 sub-display setting
02
00:P/N 01:SP1 02:SP2 03:SP3 04:SP4
05:Pass judge count 06:Over judge count
07:Under judge count 08:Error judge
count
108
Buzzer sound (External input detection)
00
00:Buzzer sound, 01:No Buzzer sound
110
Weight Unit
00
00:kg, 01:g, 02:ton
111
Language
00
00:Korean, 01:English
201
EMPTY Range
00
00~999999
202
Auto Zero Range
00
00~99 (Unit:0.25gradation)
203
Steady Range
08
01~99 (Unit:0.25gradation)
204
Steady condition check time
10
01~99 (Unit:0.1sec.)
205
Digital Filter
20
01:Weak vibration ~ 99:Strong vibration
206
Zero key operation mode
00
00:Always active
01:Active under steady condition only
207
Tare Key operation mode
00
00:Always active
01:Active under steady condition only