- 12 -
SI 560A DIN SIZE WEIGHING INDICATOR
5. SET-UP
5-1. Test Weight Calibration Mode (Using test weight)
5-1-1. Start Test Weight Calibration Mode
5-1-2. Setting “Capacity of weighing Scale”
If you want to set Max capacity as 1,000kg and the division is
0.1 (100g), then just input “1000”.
Remove “CAL-BOLT” on the Front panel, .
and press “CAL - LOCK S/W” inside.
When “CALIBR” displays, press
key,
select
“WCAL”
and press
key.
After displaying “CAPA”, input max capacity with keys & Press
key to save &
move to next step.