Digital Weighing Indicator
SI 560A
User Manual
2014.06.25
Страница 1: ...Digital Weighing Indicator SI 560A User Manual 2014 06 25...
Страница 2: ...Installation Components 10 4 3 Load Cell Installation 11 5 Set up 12 5 1 TEST Weight Calibration Mode 12 5 2 Simulating Calibration Mode 16 5 3 F FUNCTION Setting 20 5 4 Test Mode 31 6 Interface 32 6...
Страница 3: ...f wrongly used Copy Rights 1 All Right and Authority for this Manual is belonged to SEWHA CNM CO LTD 2 Any kinds of copy or distribution without permission of SEWHA CNM CO LTD will be prohibited 3 Thi...
Страница 4: ...ce where high voltage or heavy electric noise condition 4 When you connect with other devices please turn off the power of item 5 Avoid from water damage 6 For the improvement of function or performan...
Страница 5: ...40 14 104 Operation Humidity Range 40 85 RH Non condensing Function Calibration Mode Test Weight Calibration Mode Simulation Calibration Mode Display 6 digit 15mm 0 56inch Red Color FND 10 digit 13 5...
Страница 6: ...lay Key Pad 3 2 2 State LED STEADY When the weight is STEADY ON ZERO When the current weight is ZERO ON TARE TARE function is set ON OUT1 When OUT1 Relay is operated ON OUT2 When OUT2 Relay is operate...
Страница 7: ...g at the SP value setting stage 1 Set the TARE Function 1st input TARE 2nd input TARE Reset When HOLD or weight value is ZERO then this key doesn t work 2 Moving the cursor to left 1 Set the HOLD Func...
Страница 8: ...he Grand total data GRAND total data cannot be displayed key 4times Delete Sub total Sub total data cannot be displayed key 4times Delete Grand total GRAND total data cannot be displayed At the deleti...
Страница 9: ...C SIG SIG SHIELD 5 Relay output terminal COM terminal is common Terminal O1 O2 O3 O4 OC Relay OUT 1 OUT 2 OUT 3 OUT 4 COM 6 Analogue Output terminal Option Terminal 4 20mA Option 0 10V Option 7 Serial...
Страница 10: ...10 SI 560A DIN SIZE WEIGHING INDICATOR 4 INSTALLATION 4 1 External Dimension Cutting Size 4 2 Installation Components SI 560A User Manual External Dimension Unit mm Cutting Size Unit mm...
Страница 11: ...correctly Do not weld near the load cells Indicators or other devices Load Cell Installation 1 You can connect Max 8pcs of same capacity Load cells at once 350 2 You have to make horizontal balance on...
Страница 12: ...1 2 Setting Capacity of weighing Scale If you want to set Max capacity as 1 000kg and the division is 0 1 100g then just input 1000 Remove CAL BOLT on the Front panel and press CAL LOCK S W inside Wh...
Страница 13: ...city value cannot be over 1 20 000 If this condition is not fulfilled err 1 will be displayed and move back to capacity setting mode 5 1 4 Measuring the DEAD Weight of Weighing Scale After DIVI is dis...
Страница 14: ...D value will not be calculated Please remove the cause of the force or vibration and process it again 5 1 5 Calculating span value Indicator will search DEAE weight during 10secs automatically to find...
Страница 15: ...I 560A DIN SIZE WEIGHING INDICATOR After calculation span value will be displayed on the display Then press key When CALEND is displayed and calibration is completed Calculate Span value during 10 20...
Страница 16: ...ion mode uses Load cells max capacity and Max Output Rate mV so the weight adjustment degree might be less than Test weight Calibration The guaranteed resolution of this Simulation Calibration is 1 3...
Страница 17: ...s of load cells and each load cell s Max capacity is1 000kg Then total Max Capacity will be 4 000kg 1 000 x 4 and you have to input 4 000 5 2 3 Setting Digit Division value After CAPA displayed Check...
Страница 18: ...tput Voltage mV When DEAD displays Press key then indicator will calculate dead weight of scale part automatically While this process there should be nothing on the scale part After mV displayed Check...
Страница 19: ...rate and Real Output rate of load cell there might be some weight difference after finishing calibration If you want to make more precise weighing process please measure real output rate of load cell...
Страница 20: ...3 F FUNCTION Setting 5 3 1 Starting F FUNCTION Mode Press key 4 times When SETUP is displayed press key Input the set value and press key Whenever press key function No is increased Enter the functio...
Страница 21: ...setting 01 00 P N 01 SP1 02 SP2 03 SP3 04 SP4 05 Pass Count 06 Over Count 07 Under Count 08 Error Count 107 No 2 sub display setting 02 00 P N 01 SP1 02 SP2 03 SP3 04 SP4 05 Pass judge count 06 Over j...
Страница 22: ...ual 01 Auto at empty range 02 Auto at steady condition 03 Auto when finish relay out is off 215 Auto Tare Removal Time 00 00 Disuse 01 09 Use Unit 1 sec 216 Hold Mode 00 00 Sample Hold 01 Peak Hold 02...
Страница 23: ...al Print 234 External Input 2 Setting 04 00 Disuse 01 Zero 02 Tare 03 Tare removal 04 Tare Tare removal 05 Hold 06 Hold removal 07 Hold Hold removal 08 Start Packer Mode 09 Stop Packer Mode 10 Start S...
Страница 24: ...t Non 01 Databit 8 Stopbit 1 Paritybit Odd 02 Databit 8 Stopbit 1 Paritybit Even 03 Databit 7 Stopbit 1 Paritybit Odd 04 Databit 7 Stopbit 1 Paritybit Even 302 Serial Communication Speed 02 00 2 400bp...
Страница 25: ...Stream Mode Serial Port2 00 00 Format1 01 Format2 02 Format3 03 Format4 04 Format5 313 Date transference under stream mode Serial Port2 00 00 Countinuously 01 Single time on every steady state 02 Sin...
Страница 26: ...every steady states Recent stable weight Steady weight Steady weight 02 Auto At the first steady states Recent stable weight Steady weight Steady weight 04 Manual Auto At every steady states Current...
Страница 27: ...DATE Y M D Check Modification H06 TIME H M S Check Modification 24Hours H07 Password Setting 4digit 1 Zero Key 2 Tare Key 3 Hold Key 4 SP1 Key 5 SP2 Key Password combination within 1 5 H08 Maximum Cap...
Страница 28: ...lay Output Delay Time Function 239 When Function 103 3 or 103 6 save the date after t1 time Relay output Relay Condition Relay Condition OUT 1 Current weight SP1 ON Current weight SP1 OFF OUT 2 Curren...
Страница 29: ...or 103 6 save the date after t1 time t4 Weight Judge Time Function 243 t5 Error relay out delay Time Function 244 t6 Error relay out Time Function 245 Relay output Relay Condition Relay Condition OUT...
Страница 30: ...When Function 103 3 or 103 6 save the date after t1 time t2 Finish Relay Output Time Function 240 Relay output Relay Condition Relay Condition OUT 1 Start input ON Current weight SP1 SP3 OFF OUT 2 Sta...
Страница 31: ...st Mode 1 Analog Value Press key 4times Analog Variation Value Press key 4 times Display Press key 4 times Key Press key 4 times Test Mode 2 External Input Press key 4 times Relay Output Press key 4 t...
Страница 32: ...OAD NT NET WEIGHT Tare is not set ST STEADY GS when setting TARE US UNSTEADY 2 Data Format2 ID Number Data Transference Refer F function 101 305 01 22byte Header1 Header2 OL OVER LOAD NT NET WEIGHT Ta...
Страница 33: ...LOAD NT NET WEIGHT Tare is not set ST STEADY GS when setting TARE US UNSTEADY Lamp display Bit 0 Bit 1 Bit 2 Bit 3 Bit 4 Bit 5 Bit 6 Bit 7 ZERO TARE Gross Weight Print HOLD 1 STEADY 1 5 Format 5 P N...
Страница 34: ...transmission data Current Weight STX ID RCWT ETX 22 byte Current data STX ID RCWD ETX 46 byte Grand total data STX ID RGRD ETX 28 byte Sub total data STX ID RSUB ETX 30 byte Weighing completion value...
Страница 35: ...te Grand total Print STX ID WGPR ETX 8 byte Delete Grand total STX ID WGTC ETX 8 byte Run STX ID WSTR ETX 8 byte Stop STX ID WSTP ETX 8 byte Date setting STX ID WDAT DATE YYMMDD ETX 14 byte Time setti...
Страница 36: ...ng th Feature Description 1 2 RO Capacity 3 2 RO None 0x00 5 2 RO Analog Value 7 2 RO Span Value 9 1 RO Division 10 1 RO Decimal point 11 2 RO Current Weight 13 2 RO Tare Weight 15 2 RO Measured Weigh...
Страница 37: ...5 6 7 Steady Zero Tare OUT1 OUT2 OUT3 OUT4 Hold 3 Error register Address 21 Length 2 0 1 2 Loadcell Error Over Load Set point Error 4 Key register Address 441 Length 1 0 1 2 3 4 5 6 7 8 9 10 11 Start...
Страница 38: ...ction can be set under Function 226 229 6 2 1 Relay specification Coiling Rating 12VDC Contact Ratings 1A 24VDC 6 2 2 Relay output internal circuit If you enter to Calibration or Test mode Relay outpu...
Страница 39: ...tput Setting mode Step3 Adjust the displaying value of indicator with keys by 9 999 to make Digital multi meter s value as minimum ex 4mA and save When the SI 460C indicate about 5 200 the Digital mul...
Страница 40: ...Step1 Connect Digital multi meter to the Indicator A out terminal Step2 Enter F function HF14 Minimum Analog Output Setting mode Step3 Adjust the displaying value of indicator with keys by 9 999 to m...
Страница 41: ...nting format is already programmed and fixed with SE7200 7300 model 6 4 1 Printing Format Using the RS 485 or 422 interface please use convertor and converts to RS 232 and connect with Serial printer...
Страница 42: ...5 3 Isolate Resistance is more than 100M Weight Value is increased regular rate but not return to Zero 1 Load cell Error 2 Load cell connection Error 1 Check Load cell connection 2 Measure Load cell...
Страница 43: ...apacity Eerr06 Amp Gain is too big Check standard weight s weight with set value If there is difference between set value and real weight please re input the value set value is too small Eerr07 Amp Ga...
Страница 44: ...icator s A D with the other indicator 4 Check the power and connection of terminal UNpa55 1 Power is ON when some materials are on weighing part Under Normal Mode if there are more than 20 loading of...
Страница 45: ...whacnm s permission Any kinds of Mal function Defection or External damage caused by operator Any kinds of Mal function Defection caused by using spare part from Non Authorized Distributor or Agent An...