![Sewha SI 410 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/sewha/si-410/si-410_user-manual_1250426018.webp)
- 18 -
SI 410 WEIGHING INDICATOR
5-2-2.
Setting “Capacity of Load Cell”
In case of multiple pieces of load cells are installed, Please make sum of each load
cell’s capacity and make setting with Max Capacity.
EX) There are 4pcs of load cells, and each load cell’s Max capacity is1,000kg.
Then, total Max Capacity will be 4,000kg(1,000 x 4) and you have to input 4,000.
5-2-3.
Setting “Digit / Division” value
After “CAPA” displayed, Check Max Capacity of Load cell, Input the Max Capacity of
Load cell. (refer the load cell label, or Test Report.) And press
key.
After “DIVI” is displayed select Decimal point with
and
key, and set the
division with
and
key. Press
key to save.