background image

OPERATION

In User Set mode the CPM and UP, DOWN keys function differently 
than MP mode.  The CPM key becomes a key used to select the digits 
of the refill and discharge time settings found at the rightmost positions 
on the display.  The UP and DOWN keys are used to adjust the digit 
value up or down.  By selecting and adjusting digits up and down a user 
can quickly set any time from 00.1 seconds to 99.9 seconds.  Sampling 
in User Set mode is the same as explained, above, for MP mode.

The MP10 does not attempt to translate a user set time into a corres-
ponding ID or CPM.  Also, any settings you have entered in MP or ID 
modes are lost once you press the MODE key to enter MN mode.  

Note: 

changes in time settings that are entered while the controller is cycling 
are reflected on the next cycle change (so a long refill time of 15 sec-
onds will time out before a new refill time becomes valid).

MP10 Battery - 

The MP10 features sophisticated power-supply circ-

uitry that optimizes battery life.  A fresh set of AA batteries will pro-
vide more than 100 hours of controller operation at normal operating 
temperatures.  As ambient temperatures drop below 15-20F (-9C to 
-6C), the ability of  the  alkaline batteries to deliver energy is affected.  
Continuous operation may be difficult in extremely cold conditions.  
Once the batteries and MP10 warm, additional cycle capacity will be 
regained from a set of batteries. drop below 15-20F (-9C to -6C), the 
ability of  the  alkaline batteries to deliver energy is affected.  Contin-
uous operation may be difficult in extremely cold conditions.  Once the 
batteries and MP10 warm, additional cycle capacity will be regained 
from a set of batteries.

Replace alkaline batteries by removing the 4 thumbscrews located on 
the battery cover and inserting 3 fresh cells.  The MP10 battery holder 
includes space for 

3 spare AA cells

 so you should never be without 

power in the field.  Properly dispose of the spent alkaline cells.

Note: If you are storing the MP10 for more than 3 months, remove the 
AA batteries to prevent leakage.  The MP10 power supply is automatic-
ally shut off by closing the lid. Make sure the lid is closed during 
storage.

 

15

Содержание QED MicroPurge basics MP10

Страница 1: ...1 800 624 2026 FAX 734 995 1170 info qedenv com www qedenv com 1133 Seventh Street Oakland CA 94607 2601 1 800 537 1767 FAX 510 444 6789 QED ENVIRONMENTAL SYSTEMS Ground Water Sampling Remediation Lea...

Страница 2: ...MP10 Controller Optional MP30 Drawdown Meter Well Cap Tubing Air Supply In Pump Air Supply Pump Discharge Well Wizard Bladder Pump R MP10 Basic Setup...

Страница 3: ...the MP10 On Opening Display MicroPurge Mode Using CPM Sampling Use with the MP30 Level Sensing Water Level Meter ID Mode User Time Set MN Mode MP10 Battery Troubleshooting MP10 Specifications QED Serv...

Страница 4: ...ng instructions before operating the MP10 controller Warning Do not disassemble the pneumatic pump while it is connec ted to a compressed gas source Dangerous pressures could cause injury Diagrams and...

Страница 5: ...MN Manual Time Set LVL LevelShutoff CYCLE I I MODE CPM Value Flow Value Flow Value Hold Sample Cycle Start Stop ID MN MP Battery Mode CPM ID Time Refill Discharge 00 1 103 10 0 MP CPM4 05 0 D D E E C...

Страница 6: ...f pump is Refilling or Discharging Cycles per minute setting The mode the controller is in Refill time setting in seconds Time remaining in the cycle ID for time settings Abbreviations CPM MP ID HELD...

Страница 7: ...de water level feedback The MP30 uses a stan dard conductivity probe to detect the ground water surface and a marked tape allowing the user to measure the depth When the meter is set in MicroPurge mod...

Страница 8: ...approximate depth of the sample pump Follow instructions on the battery panel Opening the MP10 case turns power ON Select desired Cycles Per Minute CPM with default value is 4 CPM lower CPM for deeper...

Страница 9: ...red with its lid closed Opening Display The opening display is shown for 5 seconds and displays the controller name the version number and the battery voltage as shown in Figure 1 Figure 1 shows that...

Страница 10: ...ode Figure 2 MP10 MicroPurge Mode Using CPM The MP10 introduces a revolutionary simpler way to con trol bladder pump flow rate and achieve the low flow method used by experts Up down arrow keys are us...

Страница 11: ...ayed in the lower left corner of the display indicates Micro Purge mode The default CPM setting of 4 cycles per minute is a good starting point for wells with depths from 25 100ft MicroPurge mode star...

Страница 12: ...he pump fills and waits Pressing the PAUSE key a second time causes the controller to immediately enter the Sample state Drive air is directed to the pump causing the pump to discharge its volume of l...

Страница 13: ...his allows easy adjustment of the 2 throttle giving pressures that will produce gentle non turbulent flow normally 10 20 Feet H O deeper than pump depth For traditional 2 high volume purging pressure...

Страница 14: ...ler MP Micro Purge ID ID Time Set MN Manual Time Set LVL LevelShutoff CYCLE I I MODE CPM Value Flow Value Flow Value Hold Sample Cycle Start Stop ID MN MP Battery Mode CPM ID Time Refill Discharge 00...

Страница 15: ...ig nals the MP10 to resume pump operation The MP10 resumes by start ing in the refill leg of the pump cycle The normal operating mode for using the MP30 with the MP10 is 1 Use the MP30 in standard WLM...

Страница 16: ...ditional information on the MP30 is given in the MP30 O M manual OPERATION ID Mode Figure 6 shows an example of the MP10 in ID time set mode Once you ve used the MP10 in MP mode and found proper setti...

Страница 17: ...econds will time out before a new refill time becomes valid User Set Mode A final controller mode User Set mode MN on the display is useful for manually setting refill and discharge times on the contr...

Страница 18: ...battery life A fresh set of AA batteries will pro vide more than 100 hours of controller operation at normal operating temperatures As ambient temperatures drop below 15 20F 9C to 6C the ability of th...

Страница 19: ...en in MP mode Temperature below 10F Warm controller Air not cycling through controller Throttle turned too low Turn throttle clock wise to produce pressure Air source not delivering air Verify air sou...

Страница 20: ...s intr oduced from improper battery installation or switch connections LCD display 32 character 2 lines 16 charact ers each Non glare double hardened optical acrylic Sealed NiCd battery pack and 3 use...

Страница 21: ...e defective product with a comparable new unit at no charge to the Buyer or to refund the full purchase price In no event shall such allegedly defective products be returned to QED without its consent...

Страница 22: ...e If the return of a component is determined to be necessary QED will authorize the return of the component at owner s expense If the product proves not be defective within the terms of this warranty...

Страница 23: ...13 5 13 12 5 12 11 5 11 10 5 9 5 9 8 5 8 7 5 7 6 5 6 5 5 5 10 30 31 32 33 34 35 36 37 38 39 40 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38...

Отзывы: