In the event of fire the end user must ensure that the system pressure is vented to a safe
area.

*

The effects of decomposition of unstable fluids should be considered by the user when
placing this device in service.

*

The pressure transducer has no means of draining or venting, this must be performed by
another component in the end users system.

*

Pressure range must be compatible with the maximum pressure being measured.

*

Pressure media must be compatible with the transducer wetted parts which are:-

Ranges 1 and 1.6 bar, 15 and 30 psi

 :

15-7 Mo plus 17-4 PH stainless steels

Ranges 2.5 bar and 60 psi and above

 :

17-4 PH stainless steel

Pressure connection code B2 

:

Inconel alloy 625

Ranges 9000 and 10,000 psi:

Inconel alloy 625

*

Exposed end of cable must be kept free from moisture.

*

Liquid must not be allowed to freeze in the pressure port.

MECHANICAL INSTALLATION

Pressure Connections

:  G¼ internal pressure connection as standard.  Alternative fitted as specified

at time of order.

Pressure couplings screwed into G¼ pressure ports should have a maximum thread engagement ½”
(13mm) and 

UNDER NO CIRCUMSTANCES

 be allowed to touch the pressure sensitive

diaphragm.  Pressure couplings should be sealed against the outer face at the pressure port entry using
bonded seal washer such as:-

Part

Number

Description

Dowty Ref

GA29

G¼ bonded seal up to +100°C operation

400-021-4490-02

GA30

G¼ bonded seal up to +200°C operation

300-021-0967-02

GA31

G1/8 bonded seal up to +200°C operation

300-020-0967-02

Refer to Dowty for external pressure ratings.

Mounting

: Pressure Transducer is designed to be attached by the coupling thread only. Omni-

directional.  To fit, use a 7/8 inch AF (22.2mm) wrench on the hexagon provided and apply maximum
torque of 20  lbf-ft (27Nm). The Customer must ensure that the pressure seal is suitable for the
application.  If in doubt contact Setra Systems.

Vibration

:  Where present, mount in a saddle clamp such as part number  GA17 (material:

Polypropylene) or  GA20 (material: Polyamide).  Position the clamp as close to the pressure port as
practicable.

Effects of Heat

:  Avoid mounting the transducer near a source of heat which is liable to create a

temperature gradient across the instrument.  If this is unavoidable, use a heat shield to deflect uneven
radiated heat or wrap the transducer in glass fibre insulation so that an even temperature is assumed
throughout.

Содержание 540 Series

Страница 1: ...Installation Guide MODELS 540 541 542 Pressure Measuring Transducers 560550 0085 Rev A 1 800 257 3872 TOLL FREE 1 978 264 0292 FAX www Setra com Web Site ...

Страница 2: ...Setra offers a complete line of products for these industries Industrial HVAC Test Measurement Barometric Ultra High Purity Sanitary ...

Страница 3: ...LE section All instruments conform to the appropriate specifications and or drawings applicable and have been subjected to relevant strict quality control procedures HAZARDOUS PRODUCTS The Consumer Protection Act of 1987 Section 6 of the Heath and Safety at Work Act 1974 and the Control of Substances Hazardous to Health Regulations 1988 require that we advise recipients and users of our products o...

Страница 4: ...UNDER NO CIRCUMSTANCES be allowed to touch the pressure sensitive diaphragm Pressure couplings should be sealed against the outer face at the pressure port entry using bonded seal washer such as Part Number Description Dowty Ref GA29 G bonded seal up to 100 C operation 400 021 4490 02 GA30 G bonded seal up to 200 C operation 300 021 0967 02 GA31 G1 8 bonded seal up to 200 C operation 300 020 0967 ...

Страница 5: ...ression devices providing surge immunity to IEC 801 Part 5 To protect the receiving equipment it is recommended the end user fits a Lightning Protection Unit LPU at the instrumentation end close to the point of cable entry into the building installed as recommended by the LPU manufacturer to protect against the likely high current surge resulting from lightning strikes Types 540 542 Weatherproof C...

Страница 6: ...e CALIBRATION Transducers are calibrated to the datum requested at time of order this can be identified as follows Code G gauge datum vented to atmosphere via the electrical connection Code A absolute datum Code S sealed reference reference side of the instrument is sealed and the output electrically adjusted to zero with 1013 mb applied to pressure port Code U uni directional differential Code B ...

Страница 7: ...n 50V d c for this test Causes of Failure Excess pressure of more than twice the rated range 1 5 times rated range for Inconel wetted parts versions A short duration pressure cycle having such a peak value may be sufficient to damage the strain gauge The instrument may withstand some excess pressure without failure but a change in residual output may result If this is not excessive the instrument ...

Страница 8: ...Setra Systems with the following 1 Name and phone number of person to contact 2 Shipping and billing information 3 Full description of malfunction 4 Remove any pressure fitting and plumbing that you may have installed and enclose any required mating electrical connectors and wiring diagrams Please ship to Setra Systems at the address shown below or to your local Setra Distributor when a replacemen...

Страница 9: ... INSTRUMENT CONTAINS VITON W A R N I N G ELECTRICAL REQUIREMENTS SUPPLY VOLTS MODEL E LECTRI CAL C ODE MAX RATED OUTPUT ve SUPPLY ve SUPPLY ve OUTPUT ve OUTPUT CASE EARTH 540 541 5 42 B3 Pin A Pin D E Pin B Pin C F 541 542 N4 Red White Yellow Blue Shield 541 542 UA Red White Yellow Blue Shield 541 B4 15V d c 10V d c 0 30mV at rated supply Pin 1 Pin 4 Pin 2 Pin 3 Pin 5 B3 10 6 Bayonet Connector N4 ...

Отзывы: