background image

CAUTION:

The 280G Series is designed to have current flow in one direction only — please observe polarity.  

We suggest that the electrical cable shield be connected to the system's loop circuit ground, thereby 

improving electrical noise rejection.

Excitation

• Minimum supply voltage (VDC) = 9 + 0.02 x (resistance of receiver plus line).

• Maximum supply voltage (VDC) = 30 + 0.004 x (resistance of receiver plus line).

• Calibrated at factory using a 250 ohm load at 24 VDC.

• Variations in the power supply voltage cause less than 0.02 mA change in the transmitter’s  

current output, per volt change in the power supply.

Calibration

Every model 280G is factory calibrated and should require no field adjustment. Whenever possible, 

any zero and/or span offsets should be corrected by software adjustments in the user’s control sys-

tem. However, both zero and span adjustments are accessible by removing the pan head screws and 

turning the zero and span potentiometer screws inside. Pan Head screws should be replaced after 

adjustment to maintain enclosure rating. Voltage output units are factory calibrated into a 50 K load. 

Voltage units are operable into a 5000 load or greater. The zero factory setting is ±50 mV. Current 

outputs units (4-20 mA) are factory calibrated with a 24 VDC loop to supply voltage and a 250 load. 

The zero factory setting is a ±0.08 mA. The span (full scale) factory settings is ±0.16 mA. Zero and 

span adjustment are approximately ±1.0 mA.

Returning products for repair

Please contact Setra (1-800-257-3872, 978-263-1400) before returning unit for repair to review in-

formation relative to your application.  Many times, only minor field adjustments may be necessary.  

When returning a product to Setra, the material should be carefully packaged and shipped prepaid to:

Setra Systems, Inc.

159 Swanson Road

Boxborough, MA 01719

Attn:  Repair Department

To assure prompt handling, please supply the following information and include it inside the package 

of returned material:

   

 

1.  Name and phone number of person to contact.

2.  Shipping and billing instructions.

3.  Full description of the malfunction.

4.  Identify any hazardous material used with product.

NOTES:

Please remove any pressure fittings and plumbing that you have installed and enclose any required 

mating electrical connectors and wiring diagrams.
Allow approximately 3 weeks after receipt at Setra for the repair and return of the unit.  Non-warranty 

repairs will not be made without approval and a purchase order to cover repair charges.

Calibration Services

Setra maintains a complete calibration facility that is traceable to the National Institute of Standards 

& Technology (NIST).  If you would like to recalibrate or recertify your Setra pressure transducers or 

transmitters, please call our Repair Department at 1-800-257-3872 (978-263-1400) for scheduling, 

cost and turnaround estimates.