SERVOTOUGH LaserExact Monitor, Operator Manual 02950001A_1
38
C
HAPTER
4
measurements for the gas, which should be calibrated proportionally, must be
switched OFF in
Gas specific parameters
menu (only for the program running in
Advanced mode). It can be switched back ON after the calibration. Calibration
mode is not applicable to the spectral temperature.
4.3.8
TCP/IP and modem configuration
TCP/IP parameters of the embedded Ethernet board (optional) and the initialising
string of a modem can be configured using this menu.
Modem init string
This string is used to initialise the modem connected to the instrument via a null
modem cable (see Section 3.6). The string is sent to the modem during start-up of
the instrument. If the string has been modified, the instrument should be restarted
in order to re-initialise the connected modem. Normally the string should be the
same as in the example above: “AT &F0 S0=1 &W0 &Y0 O0”. If it is not, modify it
and update the instrument.
IP address
The address assigned to the Ethernet board of the instrument. Default IP address
is typically set to 192.168.1.237 for LAN communication. The user can assign any
free local IP address or a valid global IP address for direct internet connection.
IP netmask
Defines the subnet mask for LAN connection. Default netmask is 255.255.255.0.
Port number
Specifies the port number for TCP/IP communication. Default port number is 5001,
but the user can assign any free port number.
Содержание SERVOTOUGH LaserExact 2950
Страница 2: ...This page intentionally blank...
Страница 6: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 IV This page intentionally blank...
Страница 69: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 APPENDIX 63...
Страница 70: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 64 APPENDIX...
Страница 71: ...SERVOTOUGH LaserExact Monitor Operator Manual 02950001A_1 APPENDIX 65...