2.15
2.14.2
01440D1STD
1.
Follow the plumbing flow schematic shown in Figure 2.6.
2.
Leak test by connecting together the sample 'IN' and sample 'OUT'
gas connections and connect a water manometer to them.
3.
Pressurise the system to approx. 500mm water gauge and leave to
stabilise. The leak rate should be less than 2mm/min. after the level
has initially stabilised.
4.
If a leak exists, check connections using a soap solution. This testing
should also include the oxygen cell and its window.
5.
Rectify leaks and re-check.
Figure 2.6
01440D1STD Plumbing Schematic
CELL
FLOW
SENSOR
OUT
A.F.C.D
IN
OUT
BACK
PRESSURE
REGULATOR
(OPTIONAL)
RESTRICTOR REQUIRED
FOR BACK PRESSURE
REGULATOR OPTION
Содержание 01440002D
Страница 2: ...This page intentionally blank...
Страница 3: ...01440D1 Light Industrial Gas Analyser Service Manual Ref 01440 002D 5 Order as part No 01440002D...
Страница 4: ...This page intentionally blank...
Страница 6: ...Table of Contents 01440D Light Industrial Gas Analyser Service Manual ii...
Страница 8: ...List of Tables 01440D Light Industrial Gas Analyser Service Manual iv...
Страница 10: ...List of Figures 01440D Light Industrial Gas Analyser Service Manual vi...
Страница 32: ...01440D Light Industrial Gas Analyser Service Manual 2 20 Figure 2 11 Circuit Diagram Front Panel PCB 01420904B...
Страница 33: ...01440D Light Industrial Gas Analyser Service Manual 2 21 Figure 2 12 Circuit Diagram Front Panel PCB 01420904B...
Страница 34: ...01440D Light Industrial Gas Analyser Service Manual 2 22 Figure 2 13 Output Connections...