01440D Light Industrial Gas Analyser Service Manual.
2.12
2.10
Replacing the Front User Interface PCB (01420904)
1.
Remove the main control PCB, refer to section 2.9.
2.
Disconnect the display ribbon cable.
3.
Undo the five screws which retain the PCB.
4.
Before installing new control PCB, transfer the switch settings for the display
(reference Table 2.2 and Figure 2.4).
5.
Replace the front panel PCB, paying particular attention to the alignment of the
LED's associated with the function switches and bezels. Before tightening the
fixing screws, ensure that the switches operate without ‘sticking'.
6.
Installation is the reverse of the removal process. Ensure spacers are
transferred from old PCB as required.
Table 2.2
SW3 Switch Settings
Measurement
SW3 Settings
1
2
3
4,5,6
7
8
100% O
2
, CO
2
& CH
4
ON
OFF
OFF
OFF
OFF
OFF
50% CO
2
, CO & CH
4
ON
OFF
OFF
OFF
ON
OFF
25% CO
2
, CO & CH
4
ON
OFF
OFF
OFF
OFF
ON
10% CO
2
& CO
OFF
ON
OFF
OFF
OFF
OFF
5% CO
2
& CH
4
OFF
ON
OFF
OFF
ON
OFF
2.5% CO
2
& CO
OFF
ON
OFF
OFF
OFF
ON
1% CO
2
& CO
OFF
OFF
ON
OFF
OFF
OFF
0.5% CO
2
OFF
OFF
ON
OFF
ON
OFF
0.25% CO
2
OFF
OFF
ON
OFF
OFF
ON
Содержание 01440002D
Страница 2: ...This page intentionally blank...
Страница 3: ...01440D1 Light Industrial Gas Analyser Service Manual Ref 01440 002D 5 Order as part No 01440002D...
Страница 4: ...This page intentionally blank...
Страница 6: ...Table of Contents 01440D Light Industrial Gas Analyser Service Manual ii...
Страница 8: ...List of Tables 01440D Light Industrial Gas Analyser Service Manual iv...
Страница 10: ...List of Figures 01440D Light Industrial Gas Analyser Service Manual vi...
Страница 32: ...01440D Light Industrial Gas Analyser Service Manual 2 20 Figure 2 11 Circuit Diagram Front Panel PCB 01420904B...
Страница 33: ...01440D Light Industrial Gas Analyser Service Manual 2 21 Figure 2 12 Circuit Diagram Front Panel PCB 01420904B...
Страница 34: ...01440D Light Industrial Gas Analyser Service Manual 2 22 Figure 2 13 Output Connections...