Table 1: Specifications 

 

Volume inner buffer chamber 

200 ml 

Volume outer buffer chamber 

450 ml 

Voltage (max) 

500 V 

Current (max) 

250 mA 

Operating temperature 

4 °C – 65 °C 

Electrodes 

Platinum wire (0.2 mm, 99.99 % pure) 

Dimensions 

16 x 15.6 x 9.5 cm (WxHxD) 

Weight 

1.2 kg 

 
 

3.  Operational conditions 

 

This apparatus is intended for indoor use only. Maximum relative humidity up to 
80 % (for temperatures up to 31 °C) decreasing linearly to 50 % relative humidity 
(for temperatures up to 40 °C), at maximum altitude of 2000 m (MSL). 
 

 

4.  Using the vertical gel electrophoresis unit 

4.1. Safety precautions 

•  

Read

 the instructions carefully before using the apparatus. 

 

•   Always disconnect the electrophoresis unit from the power supply before removing 

the safety cover. 

 

•  

Do not

 attach the safety lid to the internal gel running unit while it is out of the 

bottom buffer chamber. 

 

•  

Do not

 exceed the maximum operating voltage or current (see Table 1). 

 

•  

Do not

 operate the electrophoresis units in metal trays.  

 

•   Polymerised gels may contain some unpolymerised acrylamide monomer, which is 

a volatile, cumulative neurotoxin and suspected carcinogen. Wear effective 
protective clothing and gloves. Follow recommended handling and disposal 
procedures. 

 

•  

Do not

 fill the unit with running buffer above the maximum fill lines. 

 

•  

Do not

 move the unit when it is running. 

 

•  

Caution:

 During electrophoresis very low quantities of various gases are produced 

at the electrodes. The type of gas produced depends on the composition of the 
buffer employed. To disperse these gases make sure that the apparatus is run in 
a well ventilated area. 

 

DOMINIQUE DUTSCHER SAS

Содержание BlueVertical PRIME

Страница 1: ...STRUCTION MANUAL BlueVertical PRiME Vertical Electrophoresis Chamber SERVA Electrophoresis GmbH Carl Benz Str 7 D 69115 Heidelberg Phone 49 6221 138400 Fax 49 6221 1384010 e mail info serva de http ww...

Страница 2: ...hat your expectations are reflected in the successful use of the apparatus First check that the apparatus has been received complete and undamaged following shipment Any faults or losses must be notif...

Страница 3: ...ral care and maintenance 6 4 3 Gel and buffer volumes running conditions 6 4 4 Performing electrophoresis 7 4 4 1 Insertion of pre cast gel cassettes into the inner core running unit 7 4 4 2 Sample lo...

Страница 4: ...nstruction All acrylic joints chemically bonded Doubly insulated cables rated safe up to 1000 volt Gold plated electrical connectors corrosion free and rated safe up to 1000 volt Safety lid with integ...

Страница 5: ...removing the safety cover Do not attach the safety lid to the internal gel running unit while it is out of the bottom buffer chamber Do not exceed the maximum operating voltage or current see Table 1...

Страница 6: ...ble1 The running conditions Table 2 vary according to the number of gels and their composition The current required will increase in proportion to the number of gels providing that the voltage is not...

Страница 7: ...he precast gel cassette into the inner core running unit When running only one gel a dummy plate is required on the opposite side to retain the required buffer level Place the gel running unit into th...

Страница 8: ...ctrophoresis Close the safety lid firmly making sure that the electrical connectors form a good contact Connect the electrophoresis apparatus to the power pack and connect the power pack to the mains...

Страница 9: ...rode connectors with tissue Ensure that the connectors are clean and dry before usage or storage see chapter 4 2 For further information about the SERVA BlueLine please contact the technical service o...

Страница 10: ...43277 SERVAGel TG PRiME 8 16 Tris Glycine Gel 8 16 12 43279 10 43280 2D 43281 SERVAGel TG 8 Tris Glycine Gel 8 Available upon request please inquire 12 43208 10 43209 SERVAGel TG 10 Tris Glycine Gel...

Страница 11: ...DS Electrophoresis Buffer 20x 42560 Buffers for native PAGE SERVA Tris Glycine Native Sample Buffer 2x 42528 SERVA Tris Glycine Native Electrophoresis Buffer 10x 42530 Native Anode Buffer for BN CN 10...

Отзывы:

Похожие инструкции для BlueVertical PRIME