4
LISST-Portable|XR manual
Outputs: Total volume concentration, mean size, standard deviation, optical transmission, D5, D10, D16, D25, D50
(median grain size), D60, D75, D84, D90, D95, D60/D10 (Hazen uniformity coefficient), particle surface area, silt
fraction, silt volume, size distribution, battery voltage, sample notes.
Resistant to Ethanol, IPA, diesel.
Optional built-in ultrasonic probe for complete particle dispersion.
Small-angle forward laser light scattering (laser diffraction), compliant with the ISO-13320-1 standard.
2.3 Specifications
2.3.1
SIZE AND CONCENTRATION RANGE
0.34-
500 μm in 44 size classes.
30-1,900 mg/l. Concentration range depends on the size distribution of the particles, as per examples in the table
below.
Concentration resolution < 1 mg/l.
Concentration accuracy: ± 20%.
Material
Concentration [mg/l] @
95% transmission
Concentration [mg/l] @
75% transmission
D10
[µm]
D50
[µm]
D90
[µm]
SMD
[µm]
ISO Fine
30
170
1.5
7
41
3
ISO Coarse
95
395
4
38
99
10
20-30 µm glass beads
195
1075
19
24
34
24
Sieved sand 75-125 µm
345
1925
85
122
175
112
2.3.2
MECHANICAL AND ELECTRICAL
Dimensions and weight of instrument: 17.7 cm (7”)
29 cm (11.5”)
44.3 cm (17.5”) [H
D
W]; 7.5 kg (17 lbs).
Dimensions and weight in shipping box: 78 cm (31”)
53 cm (21”)
28 cm (11”); 22 kg (49 lbs).
Data storage: 128 MB flash card, capable of storing at least 40,000 size distributions and associated sample
information.
RS232 interface with built-in USB-serial adapter.
Rechargeable Lithium-ion battery provides for 6 hours of sample processing
AC battery charger (120/240 VAC). The battery charger cannot overcharge the battery pack.
Optional 25W, 40kHz ultrasonic probe with controller electronics, managed from the touch panel display.
2.4 ASCII data file format (.ASC extension)
The .ASC file is a comma-separated file with 3 rows of text and info before the measurement data. If single sample was chosen
as storage mode (see chapter
Error! Reference source not found.
) only one additional row (#6) will be present. If multiple
amples was chosen as storage mode several additional rows, one per sample, will be present in the file.
Row 1
Header identifying file as a LISST-P
ORTABLE
|XR data file and the date, time and day number of file
creation.
Row 2
Instrument serial number
Row 3
Firmware version and date
Row 4, elements 1-21
(columns A-U)
Headers used for displaying the data in the enclosed EXCEL spreadsheet.
Row 4, elements 22-65
(columns V-BM)
Midpoint of the 44 LISST-P
ORTABLE
|XR size classes in µm.
Row 4, elements 66-84
(columns BN-CF)
Headers used for displaying the data in the enclosed EXCEL spreadsheet.
Row 5
Units for the headers in row 4.
Row 6, element 1
Date in MM/DD/YYYY format
Содержание LISST-PORTABLE XR
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ...2 LISST Portable XR manual This page intentionally left blank ...
Страница 14: ...8 LISST Portable XR manual This page intentionally left blank ...
Страница 22: ...16 LISST Portable XR manual This page intentionally left blank ...
Страница 24: ...18 LISST Portable XR manual This page intentionally left blank ...
Страница 26: ...20 LISST Portable XR manual This page intentionally left blank ...
Страница 28: ...22 LISST Portable XR manual This page intentionally left blank ...
Страница 32: ...26 LISST Portable XR manual This page intentionally left blank ...
Страница 38: ...32 LISST Portable XR manual This page intentionally left blank ...