22
5.10
Exporting Recordings
Vetcorder has built-in Bluetooth wireless connectivity, which
enables exporting measured records to mobile phones and tablets
running iOS or Android platform. After downloading the
Vetcorder Connect app, follow these instructions to pair and send
recordings to your mobile phone:
1.
Tap the <
Upload
> icon, then the device will enter Bluetooth
mode, and the screen will show the Bluetooth icon in the
middle of screen.
2.
On the mobile app, select your Vetcorder to pair, then
navigate to your recording type to select and download
recordings.
Note: Full instructions available at www.sentierconnect.com
6.
Review
To open the Data Review menu
1.
On second menu screen, tap the <
Review
> icon.
2.
In Data Review, select icon of the records you would like to
view.
6.1
Reviewing Daily Check
To review Daily Check records
1.
In the Data Review menu, select:
2.
Choose the appropriate Pet or Guest to open the list as below,
and then select one record to review more information as
below.
In this menu, you can:
Select to delete this measurement.
Select to replay the ECG waveform as shown below.
23
When the ECG waveform is being replayed, you can
Select
to change the waveform amplitude.
Select to pause it.
Select to return Daily Check list.
After the ECG waveform is replayed, it will automatically return to
the previous interface.
Select to view the trend of heart rate or SpO
2
.
Select to return to the Daily Check list.
6.2
Reviewing ECG Recorder
To review ECG Recorder records, in the Data Review menu, select . The
operations you can perform are almost the same with Daily Check,
however, there is no trending graph.
6.3
Reviewing Pulse Oximeter
To review Pulse Oximeter records, in the Data Review menu, select .
The operations you can perform are the same as ECG Recorder.
6.4
Reviewing Thermometer
To review Thermometer records in the Data Review menu, select . The
operations you can perform are the same as ECG Recorder, however,
there is no waveform to replay.
24
7.
Maintenance
Warnings and Cautionary Advice
Call Sentier customer service if you think your device needs repair
Do not perform your own maintenance on the Vetcorder;
performing your own repairs will void the warranty.
7.1
Warranty
The Vetcorder is warranted to be free from manufacturer defects in
materials and workmanship within the warranty period, which starts
upon your date of purchase:
1-year warranty on the Vetcorder Monitor and Red Case.
6-month warranty on all Vetcorder Accessories.
7.2
Battery
This monitor is designed to operate on rechargeable Lithium-ion battery.
The battery charges automatically when the monitor is connected to AC
power or devices that can output electronic power through USB
connector, such as personal computer and mobile battery bank etc.
On-screen battery symbols indicate the battery status as follow:
The battery is fully charged.
The solid portion represents the remained battery energy. If the
solid portion moves from left to right, then it means that the
battery is being charged.
Indicates that the battery is almost depleted and needs to be
charged immediately. If not charged the device will shut down
automatically.
To charge the battery,
1.
Connect the smaller end of the USB charging cable to the connector
at top of monitor. The Vetcorder may not be used while charging.
32
Guidance and Declaration - Electromagnetic Immunity
The Vetcorder is intended for use in the electromagnetic environment specified below. The customer or the User of the Vetcorder
should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Voltage dips, short
Interruptions and Voltage
variations on power
supply input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Mains power quality should be that
of a typical commercial or hospital
environment. If the User of our
product requires continued operation
during power mains interruptions, it
is recommended that our product be
powered from an uninterruptible
power supply or a battery.
Power frequency (50/60
HZ) magnetic field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
33
Guidance and Declaration - Electromagnetic Immunity
The Vetcorder is intended for use in the specified electromagnetic environment. The customer or the User of the Vetcorder
should assure that it is used in such an environment as described below.
Immunity test
IEC60601 test
level
Compliance
level
Electromagnetic environment - guidance
Conduced RF
IEC61000-4-6
3 Vrms 150
kHz to
80 MHz
outside ISM
bands
3 Vrms 150
kHz to
80 MHz
outside ISM
bands
Portable and mobile RF communications equipment
should be used no closer to any part of the system,
including cables, than the recommended separation
distance calculated from the equation appropriate for the
frequency of the transmitter. Recommended separation
distances:
Radiated RF
IEC61000-4-3
3 V/m 80 MHz
to
2.5 GHz
3 V/m 80 MHz
to
2.5 GHz
Recommended separation distances:
80 MHz
㹼
800 MHz:
800MHz-2.5GHz:
Where,
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
a
, should be less than
the compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment
marked with the following
symbol:
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the
device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the device.
b
Over frequency range 150kHz to 80MHz. For Resp field strength should be less than 1V/m.
P
d
2
.
1
P
d
2
.
1
P
d
3
.
2
34
Recommended separation distances between portable and mobile RF communications equipment and the device
The Vetcorder is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the User of the Health Monitor can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the monitor as recommended below,
according to the maximum output power of the communications equipment.
Rated max. output
power of transmitter
(W)
Separation distance according to frequency of the transmitter (m)
150 kHz - 80 MHz
80 MHz - 800 MHz
800 MHz - 2.5 GHz
0.01 0.12
0.12
0.23
0.1 0.38
0.38
0.73
1 1.20
1.20
2.30
10 3.80
3.80
7.30
100 12.00
12.00
23.00
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
P
d
2
.
1
P
d
2
.
1
P
d
3
.
2