Symptom/Error Message
Possible Cause of Failure
Possible Solution
No last response code given
Communication problem between SoloPORTER
and probe
Check probe cabling, interface and all connections
No status lights when SoloPORTER is first
plugged in
Both SoloPORTER and Solo Head Unit batteries
are too low
Replace SoloPORTER and Solo Head Unit batteries
Red light blinks whilst green light is solid
Low Solo Head Unit batteries
Replace Solo Head Unit batteries
Red light blinks after green light goes off
Low SoloPORTER battery
Replace SoloPORTER battery
No USB flash drive in SoloPORTER
Insert USB flash drive and confirm is fully inserted into
socket
USB flash drive is write protected
Unprotect the flash drive or replace with unprotected flash
drive
Wires disconnected from probe
Make sure wires are configured correctly and not loose
Wires connected to wrong position
Make sure wires are configured correctly and not loose
Probe not configured correctly
Check probe configuration
A battery cap taken off (no Head Unit power)
Either a battery cap taken off or not sealing correctly
Solid red for 5 seconds, followed by solid green
and flashing red for 8 seconds
Incorrect firmware on probe or SoloPORTER
Probe requires firmware version 1.2.2 (or later) and
SoloPORTER requires “VDAP0364.FTD” file (or later) in
the SoloDATA directory