System Setup
Field Installation
Copyright © 2008-2012 Sentek Pty Ltd All rights reserved
Page 10
Power tab
1. Select
Power
tab to show the battery voltage supplied to the probe. The probe operating thresholds should
only be changed after advice from Sentek. The probe supply voltage should be
≈
6.1V with new batteries.
Figure 8: Power tab
WARNING:
The
Disable Probe
voltage value should always be 4 volts or higher
Sensor Test tab
1. Within the Sensor Test tab check sensors are working and reporting values by selecting “
Query All Sensors
”.
When the probe is correctly configured and installed there should be no INVALID readings for “Calibrated
Value” of individual sensors (ignore the “Total” row).
2. To stop sensor test click “
Stop Sensor Querying
”
3. Make sure all changes have been written to probe
4. Back up configuration by clicking on
Backup Configuration
, and saving to a desired location.
F
F
I
I
E
E
L
L
D
D
I
I
N
N
S
S
T
T
A
A
L
L
L
L
A
A
T
T
I
I
O
O
N
N
1. Install access tubes and mounting poles per the Sentek installation manuals (EnviroSCAN or EasyAG).
2. Secure Solo Head Unit to mounting pole or post using mounting brackets
3. Place ferrite bead on grey data cable
4. Insert the grey data cable through the cable gland on top cap into the access tube, and refit ferrite bead as
close as possible to the probe
5. Install the probe with the sensors configured as required. (See Probe Configuration Utility manual for
configuration information)
6. Connect the Solo Head Unit cable to the green Phoenix connector at the top of the probe interface in order of
colours shown in Figure 2 and Figure 3.
7. Connect to probe via Solo Download Cable or PConfig cable and open PConfig on your computer.
a) Within the
Clock
tab check that the date and time are correct.