1.3 TCSMA System Specifications
Measuring Range
Manual selection of one of the following ranges,
500 to 2,500 uS/cm
2,500 to 15,000 uS/cm
15,000 to 100,000 uS/cm
100,000 to 1,000,000 uS/cm
Enclosure
NEMA 4X, IP66, polycarbonate, stainless steel
hardware, weatherproof and corrosion resistant,
HWD: 4.4" (112 mm) x 4.4" (112 mm) x 3.5" (89
mm)
Mounting Options
Wall, panel, pipe, DIN rail, or integral-sensor.
Conduit Openings
Standard: 1 - PG-9 opening, 1 - 1" NPT center
opening, cordgrips and PG-11 adapter included.
Weight
1 lb. (0.45 kg)
Ambient Temperature
Transmitter Service, -20 to 60 °C (-4 to 140 ºF)
Sensor Service, -5 to 110°C (23 to 230 °F)**
Storage, -30 to 70 °C (-22 to 158 ºF)
Ambient Humidity
0 to 95%, non-condensing
Location
Designed for hazardous and non-hazardous areas
EMI/RFI Influence
Designed to EN 61326-1
Output Isolation
Inherently isolated by sensor design
Temperature Input
Pt1000 RTD with automatic compensation,
compensation adjustable from 0.0 to 4.0%/°C.
Sensor
Noryl body toroidal, 150 psig max, ¾” NPT rear
process connection, P1000 RTD, CPVC 20 ft
length integral cable.
Max. Sensor-to-Transmitter
Distance
200 feet (61 meters)
Power
16-35 VDC (2-wire device)
DC Cable Type
Belden twisted-pair, shielded,
Maximum length 3000 ft (914 meters)
Insertion Loss
16
VDC
** Note – sensor cable limited to 105°C max temperature when dry and 70°C max temperature if
submerged.