![Sensorex TCSMA Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/sensorex/tcsma/tcsma_operation-and-maintenance-manual_1244086011.webp)
TCSMA Conductivity System
Part 2 – Transmitter Mounting
O & M Manual
11
REV G, 7/2009
2.4 Panel
Mount
In the panel mount configuration, the entire enclosure is
panel-mounted using a special optional sealing flange – as
shown in Figure 2-2. The panel mount kit, part number
07-0204, must be ordered separately.
The sealing flange must first be attached to the enclosure.
Fasten the flange to the rear half of the enclosure using the
four hex retainers. The flange gasket material must face
towards the rear of the enclosure. The flange is now
installed. Seal up the finished enclosure by tightening down
the four enclosure screws prior to mounting.
A special cut-out is required for this configuration, as shown
in Figure 2-2. Once the cut-out has been completed, insert
the flanged enclosure through the cut-out. The mounting
bracket is then attached to the rear of the enclosure as
shown. Install the four tension screws through the four
mounting holes in the bracket, and place the no-slip rubber
boots on each screw. Tighten all screws down to seal the
enclosure flange onto the panel.