17
USB Communication & Configuration Mode (Sensor)
The SPOD units allow for the reconfiguration of several parameters pertaining to
the operation of system. Adjustment of these parameters is only accessible for a
short period of time after powering on the sensor (~10s). These parameters are
stored in non-volatile memory and are retained during subsequent power cycling.
Documentation of these parameters is listed below.
Required Components:
-
SPOD Unit
-
USB data cable
-
Computer with a serial port terminal software program (e.g. CoolTerm)
Sensor Quickstart Instructions
1)
Connect the USB cable to the SPOD and computer and establish the
communication link in the terminal software.
2)
Turn on power switch and observe initialization process. After initializing the
microcontroller and printing SPOD information, the system will system will
prompt the user to:
“Enter Configuration Mode? (YES)”
3)
Configuration mode allows access to configuration settings and system
settings. To enter configuration mode type
Yes
at the prompt and hit enter.
The menus are all text-based and easy to follow. The following list contains all
the adjustable within the menu: