12

Operation and Field Use (Continued)

2.  From the IRED”, “READY FOR MEASUREMENT” display press and release the B button to  

 

  “START” a measurement.

 

    a.  The instrument will perform a zeroing process

        i.  Apply air known to be free of methane gas

        ii.  Press the B button “OK” to accept

        iii.  “SAMPLING CLEAN AIR” will be displayed with information showing the outputs  

 

          for methane (M) and light level (L) sensors. A stable reading must occur before the    

          process automatically continues.

        iv.  A cell containing methane is introduced to the light path internally as displayed by    

        “  METHANE CELL-IN”. Also displayed is information showing the outputs for    

 

          methane (M) and light level (L) sensors. A stable reading must occur before the  

 

          process automatically continues.

        v.  If the reading is not stable the zeroing process will automatically re-initiate indicated   

          by a “RE-ZEROING” message.  This process requires and addition 120-180 seconds  

          as indicated by a countdown timer.

        vi.  A stable reading is indicated by a displayed message “PLEASE BEGIN SAMPLING”

    b.  Attach the probe and place the probe in the area to be tested. Press the B button “START”   

      within the time indicated on the countdown timer.  

 

    c.  The measuring of ethane is indicated by the “MEASUREMENT” heading and the ppb/ppm    

      readings.  During a measurement, 4 consecutive 10 second readings are taken. Each  

 

      reading is averaged over the 10 second period before it is displayed.  The display will be    

      updated with the reading every 10 seconds until the 40 second period has elapsed.  The    

      final reading is indicated when the “RESULT” heading is displayed.

        i.  If the methane concentration is too high a message will be displayed indicating the    

          need to re-zero and attempt measuring again.

        ii.  For high readings move the probe slightly away from the previous source.  Ideally    

          methane concentrations less than 2000 ppm is desired.

    d.  Ethane concentration is indicated on the display showing “RESULT” and the reading.  The    

      information is also stored in the “MEASUREMENT LOG” for later download and/or review.

    e.  Pressing the A button “EXIT” will return to the main screen displaying ”IRED”, “READY FOR  

      MEASUREMENT”.

3.  Pressing and holding the A button will activate the standby mode or shut down sequence.     

 

  The display will show “SHUT DOWN” on the left side and “STAND BY” on the right.  Press the A    

  button to turn the instrument off.  Press the C button to place into standby mode to reduce battery   

  consumption.  This will also reduce the warm up time to a few minutes. Standby mode should only  

  be used for up to two hours.

Содержание IRED

Страница 1: ...instructions before use Patented SENSIT IRED INFRARED ETHANE DETECTOR 851 Transport Drive Valparaiso IN 46383 8432 MADE IN USA SENSIT Technologies is in compliance with ISO 9001 2008 Phone 888 4SENSI...

Страница 2: ...to use Always start the IRED in an area known to be gas free Tampering with this product may void the warranty Use only Sensit Technologies approved parts and accessories Never use an instrument known...

Страница 3: ...cription 4 Parts and Accessories 5 Product Specifications 6 Product Features 6 Electronic Features 7 Display Features 8 Housing Features 9 Filters 10 Field Operations 11 12 Bump Test 13 Calibration 14...

Страница 4: ...an output that is changed when certain gases absorb the light output The filter only allows specific light wavelengths to be monitored and measured The concentration of gas is proportional to the amo...

Страница 5: ...Part 873 00024 Survey Probe Filter Blue 1 Filter Part 360 00064 SENSIT IRED Calibration Kit IRED Calibration Kit Part 881 00100 Kit includes 1 Regulator 1 Cal Adapter 1 100pm Ethane Nitrogen 58L SENS...

Страница 6: ...ic Sensor Specifications IR Sensor Method Advanced Infrared Absorption Spectroscopy Detection Range 250ppb 1000ppm Warm up 30 min Calibration required 3 months Product Features The IRED is constructed...

Страница 7: ...ghting IR communication port LCD Button Operation The Sensit IRED is operated with the use of three buttons below the display They are labeled A B C The A button Activates Deactivates the instrument A...

Страница 8: ...once for manual data save GPS Connected GPS enabled not connected 8 Display Features The display provides all information including Gas concentration in PPB and PPM Battery voltage status Pump Status...

Страница 9: ...taining screws Charge Complete Green Charging Red Status Yellow Charger Connection The battery pack has LED s to indicate charge status Charging should only be performed in an area know to be free of...

Страница 10: ...r shoulder strap attachment Clips on the strap will attach securely Hydrophobic Filter Assembly WARNING Do not operate without proper filter Damage may occur to the pump and other internal parts Only...

Страница 11: ...nd Time d Warm up is indicated PLEASE WAIT with a 600 second countdown timer The countdown timer will begin after an initial warm up of 5 10 minutes e Calibration due date indicating when calibration...

Страница 12: ...imer c The measuring of ethane is indicated by the MEASUREMENT heading and the ppb ppm readings During a measurement 4 consecutive 10 second readings are taken Each reading is averaged over the 10 sec...

Страница 13: ...stable reading must occur before the process automatically continues 5 If the reading is not stable the zeroing process will automatically re initiate indicated by a RE ZEROING message This process r...

Страница 14: ...stable reading must occur before the process automatically continues 5 If the reading is not stable the zeroing process will automatically re initiate indicated by a RE ZEROING message This process r...

Страница 15: ...mplemented 10 ALARM PPM This is not implemented 11 PURGE TIME Time in Sec unit should purge system before shutting down 12 ERASE BUMP LOG Allows you to erase bump logs 13 ERASE CAL LOG Allows you to e...

Страница 16: ...Proof of purchase may be requireds before warranty is rendered Units out of warranty will be repaired for a service charge Internal repair or maintenance must be performed by a Sensit Technologies au...

Отзывы: