26
USER MENU
C
ALIBRATION
WARNING
: Using calibration kits other than recommended by SENSIT
Technologies may cause inaccurate readings. Repairs are required if
the instrument fails to calibrate. Consult the factory for details.
Calibrate the instrument using the built-in calibration cell in the carrying case.
See the calibration section in this manual for instructions on how calibrate the
instrument.
Press and release the
ON/MENU
button again to scroll to the next menu item.
Press and release the
OFF/BACK
button to exit the menu.
B
LUETOOTH
ID
Displays the Bluetooth ID number. Currently this number is not used for
anything.
S
ET
A
LARM
Password protected.
This menu allows the user to adjust the alarm threshold. By default, this is
100ppm
•m. To adjust, use the
UP
and
DOWN ARROW
buttons until the desired
number is reached. Adjustments are made in increments of 10ppm
•m,
1%LEL
•m, or 0.01%VOL•m, depending on the unit display setting. Once
correct, press and release the
MEASURE
button to save the settin
g. “OK” will
be displayed. Press and release the
ON/MENU
button to scroll to the next
menu item or press and release the
OFF/BACK
button to exit the menu.
The
GAS•TRAC
LZ-30
has 3 different alarm stages: 20%, 50%, and 100% of
the alarm threshold. The frequency of the vibration and audible alarms will
increases as the reading reaches each stage.
NOTE
: The “3 Tier Alarm” option determines whether the 20% alarm stage will
be enabled. If 3 tier is off, only the 50% and 100% stages will be enabled.