background image

1.  Install  the  batteries  by  pushing  down 

the  locking  tab  and  sliding  the  handle 

away from the top of the instrument. When 

replacing handle be sure the tab is securely 

in place.

2. 

IMPORTANT:

 CAREFULLY OBSERVE 

POLARITY WHEN CHANGING BATTERIES.  

Incorrect installation can damage an internal 

“factory service only” fuse.

3. Locate button “A” – push & hold until the 

unit powers up, then release the power button.

4.  Allow  unit  to  go  through  the  warm  up 

sequence in clean air.  At the end of warm 

up,  the  unit  will  auto  zero  and  enter  the 

working display. This requires from 40 up to 

180 seconds.

5. If 

FAIL

 is displayed for any of the sensor 

readings,  make  sure  the  instrument  is  in 

clean air; push and hold the “C” button until 

AUTO  ZERO  is  displayed.    If  this  process 

does  not  clear  the  fail  on  the  display,  this 

could indicate a problem with the instrument 

or sensor.  

6. Look at the display –CO (and optionally 

O2) readings are displayed. 

7. Attach the hot air probe assembly to the 

¼ turn twist connection on the back.

8.  Place  your  finger  over  the  inlet  of  the 

probe  and  wait  ~5  seconds  for  “FLOW 

BLOCKED” to appear on the display.  Check 

for  tight  connections  if  it  does  not  show 

“FLOW BLOCKED”.

9. You are now ready to use the instrument. 

You can now enter the area and detect gases.

10.  When  performing  CO  checks  in  high 

temp  areas  do  not  exceed  prolonged 

exposures (over 5 min) at 500°F.

11. Press and hold the “C” button to zero 

the instrument. (only in a gas free area).

12.  When  your  investigation  is  complete 

push button “A” and hold for 5 seconds until 

the instrument displays “POWER OFF” then 

release to shut off.

SENSIT

®

 CO

QUICK-START INSTRUCTIONS