Table 1 Software options for evaluating digital Mass Flow Controllers 

2.2.

 

Analog interface 

To operate a Mass Flow Controller with an analog interface (SFC54XX), you need: 

 

a tuneable power supply as setpoint source, being able to  deliver  0-5V,  0-10V  or 4-20mA,  depending on the 
version of the analog Mass Flow Controller 

 

a DB9 cable to connect the Mass Flow Controller to a 15-24V power supply 

 

a multimeter or an oscilloscope to measure the flow signal 

3.

 

Testing recommendations 

It is recommended to start by testing the communication with the device, using the EK-F5x Evaluation Kit and the Viewer 
software. The baud rate and address of the device with the viewer software can easily be changed if necessary. 

As a second step, connect the device to a gas supply (or a source of vacuum). For best accuracy, mount the Mass Flow 
Controller horizontally and use the same pressure as during factory-calibration (specified in the datasheet, by default for 
most products 3 bar inlet vs outlet). 

If the Mass Flow Controller is mounted vertically, the chimney effect can adversely impact the accuracy at low flows and 
especially at zero flow. It is possible to use Sensirion recalibration software to compensate for that effect. The 

“R

ecalibration 

software”

 can be downloaded from the Sensirion website 

and is also linked in the “Useful resources” 

Section 

8.1

In this 

case, you will have to calibrate the mass flow controller yourself in the chosen orientation. 

Try using different setpoint steps to test the Mass Flow Controller. If the chosen 

setpoint can’t be 

reached, try increasing 

the supply pressure, removing fluidic resistances like filters, or using tubes with a larger diameter. If the supply pressure 
used is significantly lower than the calibration pressure, the controller can be slower than expected. On the other hand, if 
the supply pressure is significantly larger than the calibration pressure, the Mass Flow Controller may become unstable. In 
these cases, lowering or increasing the user gain to adapt the controller accordingly. This can be done conveniently with 
the Viewer Software. For detailed instructions, see the 

Viewer Software Manual

”, linked in “Useful resources” 

Section 

8.2.1

. 

SFC5xxx/SFM5xxx Viewer software 

C/C#/Python library 

for quick evaluation 

plug-and-play software 

 

no coding required 

no custom cabling required 

for more flexible testing and prototyping 

 
 

custom coding required 

custom cabling recommended for

 

C/C#/Python 

Figure 1 Different options for evaluating digital Mass Flow Controllers 

Содержание SFC53 Series

Страница 1: ...SFM5xxx Summary This guide provides recommendations for evaluation testing and integration of Sensirion Mass Flow Controllers and Meters It is applicable to SFC55xx SFC54xx SFC53xx as well as SFM55xx...

Страница 2: ...recommendations 4 4 Common pitfalls 5 4 1 Flow appears to be off by several 5 4 2 Settling time is longer than specified 5 5 Design recommendations 5 5 1 Pressure drop and input pressure 5 6 Fluidic a...

Страница 3: ...aluate all digital Mass Flow Controllers with a DB9 plug an adapter is available for other plugs please contact Sensirion EK F5x is available from distribution Please note Mass Flow Controller is not...

Страница 4: ...ero flow It is possible to use Sensirion recalibration software to compensate for that effect The Recalibration software can be downloaded from the Sensirion website and is also linked in the Useful r...

Страница 5: ...The settling time of Sensirion s Mass Flow Controllers is specified for a step answer from 10 to 100 of full scale within 5 of setpoint When testing the settling time do not do so from a position of z...

Страница 6: ...ic manifold must be designed and manufactured by the user In Table 4 in the Useful resources Section 8 1 CAD models for a suggested manifold designs for SFC54xx SFM54xx and SFC53xx SFM53xx can be foun...

Страница 7: ...ssure 4 bar at inlet and atmospheric pressure at outlet any situation which leads to the factory calibration requiring adjustments due to the accuracy requirements e g vertical instead of horizontal i...

Страница 8: ...ass Flow Controllers SFC sensirion com products product categories mass flow controllers Mass Flow Meters SFM sensirion com products product categories gas flow sensors General technical download sect...

Страница 9: ...SFC5500 SFM5500 50sccm 0 5slm 2slm 10slm sensirion com resource cad sfc5500 SFC5400 SFM5400 sensirion com resource datasheet sfc5400 sensirion com resource cad sfc5400 sensirion com media d ocuments...

Страница 10: ...to the stability of the MEMS based sensor element and the robust mechanical design Sensirion Mass Flow Controllers do not drift and do not require recalibration in the field High manufacturing standar...

Страница 11: ...er s expense and the warranty period for any repaired or replaced product shall be limited to the unexpired portion of the original period This warranty does not apply to any equipment which has not b...

Отзывы: