DO NOT operate this product should it malfunction or require repair. Operation of a malfunctioning
product or a product requiring repair may result in property damage, serious personal injury or death.
DO NOT attempt to repair or modify this product, except as specified in the User Manual. If repair is
necessary, contact the Sensidyne Service Department to arrange for a Returned Material Authorization
(RMA) (See APPENDIX E: RETURNED MATERIAL AUTHORIZATION for details).
ALWAYS operate this product in the manner intended. Failure to do so may result in property damage,
serious personal injury or death.
Users should ALWAYS
refer to MSDS and suppliers’ instructions for proper safety instructions and
handling for any chemicals used with this equipment.
Use ONLY genuine SENSIDYNE® replacement parts when performing any maintenance procedures
described in this manual. Failure to do so may seriously impair the instrument performance. Repair or
alteration of the product beyond the scope of these maintenance instructions, or by anyone other than an
authorized SENSIDYNE® service technician, may cause the product to fail to function as designed, and
persons who rely on this product for their safety could sustain severe personal injury or death.
The SensAlarm Flex Universal Gas Monitoring System is an ambient air monitoring device. Restricting
the access of ambient air to the sensor may result in less than optimal monitoring performance.
Prolonged exposure to excessively high concentrations of Toxic gas may cause the sensor to produce
erroneous readings.
ALWAYS make use of genuine Sensidyne accessories, such as Rainshields, Splashguards, or
Calibration Adapters to protect against variations caused by environmental conditions. Note that reactive
Gases such as ETO, HF, HCN, NO2, SO2, HCl, ClO2, or Cl2 may require special considerations in use
with certain Sensidyne accessories.
ALWAYS Perform tests only within the specified operating ranges.
Sudden changes in pressure may cause temporary fluctuations in the sensor reading.
Sensors should be hot-swapped only when the Normal Operation Mode screen is displayed.
CAUTION: Failure to read and understand the User Manual may result in property damage,
severe personal injury or death.
CAUTION: Risk of electric shock if front cover is opened.
Important Calibration Considerations:
Verify the concentration of calibration gas standard before making calibration adjustments.
The concentration can be altered by:
Deterioration of the gas concentration during storage due to shelf-life.
Interaction of the calibration gas standard with incompatible materials used to contain and
transfer the gas (tubing), absorption onto or permeation through certain plastics, polymers or
elastomers.
Interaction of the calibration gas standard with ambient contaminants, such as water.
If further translation is required, please contact the Sensidyne EU Authorized Representative (see Back
Cover for contact information) or Sensidyne Customer Support at 800-451-9444 or +1-727-530-3602.