![SenseAir eSENSE Duct-TR Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/senseair/esense-duct-tr/esense-duct-tr_installation-manual_1243346005.webp)
Document
IMA0215
Rev
7
Page
5 (8)
Figure 4 PCB with jumper to configure OUT2 for current output 4-20mA or voltage output 2-10VDC
Figure 5 Enlarged picture of the PCB with the jumper set to current output (left position)
Figure 6 Enlarged picture of the PCB with the jumper set to voltage output (right position)
Self-diagnostics
The system contains a complete self-diagnostic procedure that is executed
automatically when the sensor is in operation. Sensors with display show a
wrench
if
an error is found. The wrench is shown during the first seconds after power up and if
the measuring range is exceeded. The output Out(2) indicates the same information
by setting the output voltage to 1V or 2 mA.
Option display
+
~
+
~
+
~