background image

267 Lowell Rd, Hudson NH 03051, USA      |      Phone: 1.888.328.7661     |      Fax: 1.888.329.8328      | 

   

 

www.sengenuity.com

Helping Customers Innovate, Improve & Grow

Step 3

Step 3: Setup, Locate and Calibrate Wireless Sensors

Place the sensors at designated target locations.  Figure 7 shows 

a wireless temperature sensor.  Please note the orientation of the 

antenna in the inset (picture of sensor without the cap) to correctly 

interpret Figure 8.  

Figure 7: Wireless Temperature Sensor

Figure 8: Incorrect and Correct Setup

Step 3: 

continued...

There are some important guidelines that should be followed 

when locating the sensors:

The 

 

minimum distance between two sensors must be 4 

inches (~10 cm).

The 

 

minimum distance between any given sensor and 

the interrogation antenna must be 12 inches (~31 cm).

The 

 

maximum interrogation distance between the 

interrogation antenna and the sensor is heavily influenced 

by the local RF environment. In general, that the distance 

between the interrogator and the sensor should not 

exceed 79 inches (~200 cm). 

Ensure that the sensors are placed within the lobes 

 

of the  dipole radiation pattern.  Figure 8 shows sensor 

positions that will most likely not work and also shows 

sensor positions that will work.

Ensure that the distance between Interrogator    

 

Antenna and nearest parallel metal wall >4” (~10 cm).

Содержание TempTrackr

Страница 1: ...on Software Please contact SenGenuity if any item is either missing or damaged The wireless interrogation unit needs to be powered by a 5V 5 DC power supply All temperature sensors are interrogated by...

Страница 2: ...ngenuity TempTrackr Start the Software from the Windows Start Menu or by double clicking the WSR T2 icon from the directory in which the software has been installed Figure 1 InstallShield Window for T...

Страница 3: ...ued Connect the Wireless Interrogation Unit to a power supply 5V 5 DC 350mA as shown in Figure 6 You can now connect the RS232 RS485 cable to your PC laptop Step 2 Step 2 Setup and connect wireless in...

Страница 4: ...t should be followed when locating the sensors The minimum distance between two sensors must be 4 inches 10 cm The minimum distance between any given sensor and the interrogation antenna must be 12 in...

Страница 5: ...eaders Step 3 continued Select Calibrate from the Sensors Menu to obtain the Calibration Screen Sensors are identified by model numbers that reflect nominal the frequency of each sensor Since the six...

Страница 6: ...ise the calibration will be lost if the reader is powered off Press Test VALUE_INTER again Figure 13 shows the results of a successful calibration Figure 13 Calibration procedure FCC version complete...

Страница 7: ...nd found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a reside...

Страница 8: ...pas l intensit n cessaire l tablissement d une communication satisfaisante Le pr sent metteur radio identifier le dispositif par son num ro de certification ou son num ro de mod le s il fait partie du...

Отзывы: