Maintenance and cleaning
18
Temperature: 2-Point-Calibration
To calibrate the temperature signal of the O2S, two
calibration points are necessary. The calibration point 1 must
have a lower temperature (e.g. 0°C) than calibration point 2.
This temperature must be lower than 20°C.
The calibration point 2 (higher temperature) must be
between 30°C and 50°C.
Submerge the sensor in a water bath with the lower
temperature (calibration point 1) and wait, until the sensor
has the same temperature as the water. Select in the menu
“Sensors / O2S / Sensor Calibration / Temperature /
Calibration point 1”. Measure the water temperature with a
reference thermometer and enter this temperature with the
cursor keypad. Confirm the calibration point with OK.
Repeat this procedure for the upper calibration point 2 with
a temperature between 30°C and 50°C.
Tipp: You find the calibration information in the menu under
“Sensors / O2S / Sensor Calibration / Info”.