B. QUICK GUIDE
1.
TO CALIBRATE FLAT SHEET PROBE/FORMED CAN PROBE.
A)
Attach crocodile clip to test sample but make sure probe is not
in contact
.
B)
Enter set-up mode by: pressing
CALIBRATE
and
LACQUER
NUMBER
together. (
PASS
should appear in large display, if
not press
END
and try set-up mode again) then enter
PASSWORD
. (
9963
if unchanged).
C)
Press
CALIBRATE
. (
CP
will appear in the small display).
D)
Press
ENTER
. (The large display will show
0.00
)
E)
Press
END
twice.
2.
TO SET-UP A LACQUER TYPE.
A)
Enter set-up mode by: pressing
CALIBRATE
and
LACQUER
NUMBER
together. (
PASS
should appear in the large display,
if not press
END
and try set-up mode again) then enter
PASSWORD
. (
9963
if unchanged)
B)
Press
LACQUER NUMBER
. (
??
will appear in the small
display).
C)
Enter new lacquer number by using
NUMBER KEYS
.
D)
Press
ENTER
. (
????
will appear on large display).
E)
Enter set piece value: by using
NUMBER KEYS
.
F)
Press
ENTER
.
G)
Press
END
.
3.
TO CALIBRATE A LACQUER TYPE.
A)
Press
CALIBRATE
.
B)
Press
LACQUER NUMBER
. (
??
will appear in the small
display).
C)
Enter new lacquer number by using
NUMBER KEYS
.
D)
Press
ENTER
.
E)
Place probe on known test sample. (remember to apply
pressure if the flat sheet probe is being used). Press
ENTER
before releasing.
F)
Press
END
.
4.
TO TAKE READINGS.
A)
Select appropriate lacquer number by using the
UP/DOWN
ARROW KEY
.
B)
Place probe on sample.
C)
Note readings. (Press
ENTER
if using printer or host
computer).