S model allows digital reading and control of speed.
N model allows setting a heating level as a rate of total heating
power. To regulate the temperature of the stirred liquid in a better
way, use an electronic controller Sensoterm II as an accessory.
E model allows setting the heater plate temperature. Including
1001610 probe as an accessory, it allows controlling liquid
temperature at same same time that it is stirring.
In some application it is important bearing in mind that heater plate
temperature and the liquid one are very different.
INSTALLATION
Place the stirrer on a flat, horizontal, level surface near the mains.
Connect the connection cable at the rear part of the stirrer base
and then plug it into the mains.
POWER
AGIMATIC S
25W
AGIMATIC N
550W
AGIMATIC E
Maximum temperature (N & E models)
350ºC
Temperatura resolution (E model)
± 2ºC
Heating control (N & E models)
N:0...100%
E:50...350ºC
External temperature control connection (N & E models)
Yes
Speed
60...1600rpm
Stirring power
10.000ml
Dimensions
14,5x19x26cm
Weight (S model)
3,0Kg
Weight (N & E models)
3,2Kg
Temperatura limiter (N & E models)
370...390ºC
Be sure that the voltage supply is the same as the one
indicated on the characteristics plate.
Do not use the equipment if it is not earthed.
If you change the plug, bear in mind the following:
Blue cable:
Neutral
Brown cable:
Phase
Yellow/green cable: Earth