8.7 Printer user guide
According to the requirement of customer, steam sterilizer can adopt prescriptive
printer. Connect the printer and the steam sterilizer with prescriptive cord. Switch on
the power, two lights on the printer panel will go on, and this means that the printer
has been installed.
Caution: steam sterilizer can only select micro.
Printer offered by us.
Print information explaining:
Date: Sterilize date.
CN: Name of the cycle.
C.S.T.: The starting time of the cycle.
Vp1.Vp2.Vp3: Vacuum pressure1, Vacuum pressure2, Vacuum pressure3.
Pp1.Pp2.Pp3: The phase of pressure1, the phase of pressure2, the phase of pressure3.
If you have activated the printing option at the end of the cycle, when the cycle fin-
ishes, the display will show “Ed” and at the same time “Prt” will appear, which means
that the information is sent to the printer.
8.8 Function settings
a) Press the “STEP” key for 20 seconds. Without any active program (message
“Ld”).
b) By means of the “SELECT” key you can select the alternative options:
Option 1
YEAR
“YER”
Option 2
MONTH
“MtM”
Option 3
DAY
“DAY”
Option 4
HOUR
“HOR”
Option 5
MINUTES
“MIN”
Option 6
printing
“Prt” (OFF, ON)
Option 7
EXIT
“Eit”
c) Press the “STEP” key (arrow up) or “DRY” key (arrow down) to select the
desired value.
d) Confirm the value by pressing the “START/STOP” key. The equipment will
emit a confirmation beep.
8.9 Cleaning the water tank
1) Empty the tanks water by draining it. Connect the water outlet using the one end
of the tube without connector, and the other end of the tube to a drainage sink. Then
turn the draining knob anti-clock wise to drain the water from the machine.
2) Open the tank cover board, you can use a screwdriver to loose the five screws.
3) Pull up the water tank cover by one hand.
4) Pull up the tank cover completely and then disassembly the cover.
5) Then you can start cleaning the tank. Use a cotton with some alcohol or disinfect-
ant to clean the tank. Wash it with distilled water, then empty the tank and dry it.
6) After cleaning the tank, assembly the cover and tighten the screws.
POWER THE MACHINE OFF AND LET IT COOL DOWN
BEFORE CHANGING THE SEAL
AND AVOID ANY BURNING INJURIES