0000138014
rev. 1.0 EN
5
Paragraph 3 – Proportional dosing for the Redox (O.R.P.) measurement
Programming
Operation
The pump measures and controls the Redox value of a solution, by
programming the following parameters: set-point, set-point type,
proportional band and alarm band.
Set-point type: maximum
0
mV
s/m
max
flow-rate
Alarm Band
Alarm Band
Proportional
Band
se
t-
p
o
in
t
Ala
rm
O
n
Al
ar
m
O
n
Pump Off
Pu
m
p
O
n
Set-point type: minimum
0
mV
s/m
max
flow-rate
Alarm Band
Alarm Band
Proportional
Band
se
t-
p
o
in
t
Ala
rm
O
n
Al
ar
m
O
n
Pump Off
Pu
m
p O
n
Can also be programmed:
-
The O.F.A. (Over Feed Alarm) time in minutes, a set period of time after
which, if the measured Redox value does not reach the set-point, an
alarm signal is activated.
-
Activation / deactivation of the calibration procedure
Display during operation
Allarmi e stati
•
Cal = calibrazione non
eseguita
•
Liv = Allarme livello
•
Fls = Allarme flusso
•
Alm = Misura fuori
Banda allarme
•
OFA = Allarme O.F.A.
Valore misura del
potenziale Redox.
Stato della pompa
•
Vuoto = pompa in start
•
Stop = pompa ferma
•
Paus = pompa in pausa
Valore del dosaggio in corso
(dipende dall’unità di misura
selezionata)
•
Percentuale, Frequenza,
l/h, Gph, ml/m
465mV
P 100%
PROG
Con
fi
g.
Con
fi
g.
Funct.
Funct.
Redox
Setpoint
0mV
Set. Type
High
ProBand
100mV
AlarmBnd
20mV
OFA
0m
Calibr.
Probe