10
Redox (O.R.P.) Calibration Menu
You can choose the automatic or manual mode.
-
Automatic calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press
and the
display will show the countdown of the 60 seconds required to complete the calibration. If the quality of the
alignment is less than 50%, an error appears on the display; press
to exit the calibration (after 4 seconds the
pump will automatically exit the calibration); if the quality is higher than 50%, the value appears on the display;
press
and the procedure is completed.
-
Manual calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press
and the
display will show the value 465mV flashing, insert the probe into my solution, press
to display the value of my
solution, then press
to confirm and start the calibration procedure as described previously.
PROG
Con
fi
g.
Calibr.
Probe
Calibr.
Auto.
Auto Cal
475mV
WAIT
60 s.
Error
475mV
475mV
100% ok
Calibr.
Manual
Man Cal
475mV
WAIT
60 s.
Error
475mV
475mV
100% ok