3.3.4 TEMPERATURE PROBE CALIBRATION
Connect the probe to the instrument as indicated in the electrical connections.
Select the TEMP. probe from the Calibration menu.
TEMP CAL. Type: MAN
25.0°C
CL CAL. Type: MAN
28.0°C
CL CAL. Type: MAN
28.0°C
Wait 10”
In Manual (
MAN
) mode:
•
Use a reference instrument to read the
temperature value.
•
Adjust the value shown on the display to
match the value read by the reference
instrument. Press
Enter
to confirm.
•
Wait 10 seconds for the calibration to
complete.
•
Once the operation has concluded, a
message will appear indicating that the
calibration has been carried out successfully.