0000137242 Rev.5.1
11
At the end of each calibration point, the instrument will display the quality of the electrode as a percentage.
3.3.3 CL (CHLORINE) PROBE CALIBRATION
Connect the probe to the instrument as indicated in the electrical connections.
Select the CL probe from the Calibration menu.
ORP CAL. Type: Man
ORP CAL. Type: Man
+475mV
ORP CAL. Type: Man
+475mV
Wait 60”
CL CAL. Type: MAN
0.50 ppm
CL CAL. Type: MAN
1.20 ppm
CL CAL. Type: MAN
1.20 ppm
Wait 10”
In Manual (
MAN
) mode:
•
Immerse the probe in the solution, insert the
solution’s mV value and press
Enter
•
Wait 60 seconds. When finished, the
instrument will display the probe’s quality as
a percentage value.
•
Once the operation has concluded, a
message will appear indicating that the
calibration has been carried out successfully.
MAN
In Manual (
MAN
) mode:
•
Use a reference instrument to read the
chlorine value.
•
Adjust the value shown on the display to
match the value read by the reference
instrument. Press
Enter
to confirm.
•
Wait 10 seconds for the calibration to
complete.
•
Once the operation has concluded, a
message will appear indicating that the
calibration has been carried out successfully.