Control solution testing
28
Performing a control test
⑤
Touch and hold the hanging drop of control
solution where the narrow channel meets the
TOP EDGE of the test strip. Make sure the
confirmation window fills completely. Control
solution should not be applied to the flat face of
the test strip.
⑥
Read the result on the meter
When the confirmation window is full, the meter will
count down from 5 to 1. After the measurement is
over, the meter will display the blood glucose level along
with the unit of measurement, the date and time of the test.
Results are automatically stored in the meter’s
memory.
CONFIRMATION
WINDOW
TOP
EDGE
NARROW CHANNEL
Apply Drop Here
mmol/L