14
Low-Pass Filter (50-60 Hz)
Range
Accuracy
30 to 50 Hz
50 to 60 Hz
6mA*1
±(2.0%+5D)*2
±(1.0%+5D)
60mA
±(2.0%+5D)*2
±(1.0%+5D)
600mA
±(2.0%+5D)*2
±(1.0%+5D)
6A
±(2.0%+5D)
±(1.0%+5D)
60A
±(2.0%+5D)
±(1.0%+5D)
*1 Minimum Reading is 10uA
*2 Frequency response is 15 to 50 Hz.
When frequency < 30 Hz, add 3.0% to accuracy.
Minimum Resolution :
1uA
Position Error :
±1% of reading
Overload Protection :
60A
Cut-off frequency :
200Hz
Low-Pass Filter (1k Hz)
Range
Accuracy
30 to 50 Hz
50 to 60 Hz 60 to 200 Hz
6mA*1 ±(2.0%+5D)*2 ±(1.0%+5D) ±(2.5%+5D)
60mA
±(2.0%+5D)*2 ±(1.0%+5D) ±(2.5%+5D)
600mA ±(2.0%+5D)*2 ±(1.0%+5D) ±(2.5%+5D)
6A
±(2.0%+5D)
±(1.0%+5D) ±(2.5%+5D)
60A
±(2.0%+5D)
±(1.0%+5D) ±(2.5%+5D)
*1 Minimum Reading is 10uA
*2 Frequency response is 15 to 50 Hz.
When frequency < 30 Hz, add 3.0% to
accuracy.
Minimum Resolution:
1uA
Position Error :
±1% of reading
Overload Protection :
60A
Cut-off frequency :
1kHz
Содержание MW 3950B
Страница 1: ...HFR MW 3950B Pince courant de fuite Bluetooth Notice d utilisation User s Manual...
Страница 8: ...7 Monophas Triphas Triphas Neutre Appareil Appareil Appareil...
Страница 9: ...8 Fonction HOLD MAX MIN Appuyer Appuyer 1s Appuyer Mode normal...
Страница 22: ...7...
Страница 23: ...8 Data Hold MAX MIN...
Страница 24: ...9 High Frequency Rejection 50 60 Hz 1k Hz HFR Auto Manual Range...
Страница 30: ...SEFRAM 32 rue E Martel F42100 Saint Etienne France Tel 04 77 59 01 01 Fax 04 77 57 23 23 Web www sefram com...