12
Safety Standards:
IEC / EN 61010-1
IEC / EN 61010-2-032
CAT. III 600V, Pollution Degree 2
Electromagnetic Compatibility Standards
(EMC):
EN 61326-1
Application Standards:
IEC 61557-13: Class 2, ≤ 30A/m
@ In: 3.5 to 600 mA, Fn: 40 to 1k Hz
Valid at best resolution in 6 mA, 60 mA and 600
mA range.
When measuring under the following condition
specified in IEC 61557-13, add additional
specification to accuracy specification.
Intrinsic uncertainty
or influence quantity
Designa-
tion code
Additional Spec.
Position
E1
±1% of reading
Supply voltage
E2
No additional error
Temperature
E3
0.1 x (spec. accuracy)
/ °C (<18 °C or >28 °C)
Distorted waveform
E9
No additional error
External low
frequency magnetic
field 15 Hz to 400
Hz according to
IEC 61000-4-8
E11
Add ±10µA error
per 1uT (magnetic
field)
Load current
E12
Add ±6µA error
per 1A (load current)
Touch current
caused by common
mode voltage
E13
No additional error
Frequency
E14
No additional error
Repeatability
E15
No additional error
Содержание MW 3950B
Страница 1: ...HFR MW 3950B Pince courant de fuite Bluetooth Notice d utilisation User s Manual...
Страница 8: ...7 Monophas Triphas Triphas Neutre Appareil Appareil Appareil...
Страница 9: ...8 Fonction HOLD MAX MIN Appuyer Appuyer 1s Appuyer Mode normal...
Страница 22: ...7...
Страница 23: ...8 Data Hold MAX MIN...
Страница 24: ...9 High Frequency Rejection 50 60 Hz 1k Hz HFR Auto Manual Range...
Страница 30: ...SEFRAM 32 rue E Martel F42100 Saint Etienne France Tel 04 77 59 01 01 Fax 04 77 57 23 23 Web www sefram com...