M35X3F M00
32
(12) Temperature
3535F
Function
Range
Accuracy
°C
-50 °C ~ 99.9 °C
± (1% + 2°C)
100 °C ~ 399.9 °C
400 °C ~ 1000 °C
°F
-58 °F ~ 211.9 °F
± (1% + 4°F)
212.0 °F ~ 751.9 °F
752 °F ~ 1832 °F
± (1% + 2°F)
± (1% + 1°C)
Overload protection :
1000 V
rms
-
The above specification is assumed at the ambient tempera-
ture stability within ±1°C. In addition, the temperature probe
has to be connected to meter for more than 1 hour in advance.
The meter needs 2 hour for stability for ambient temperature
change more than ± 5°C.