Disinfecting the equipment
13
Disinfecting the equipment
Please refer to the instructions for the chemical products used to comply with the correct strength
and quantity of the products, contact time, and personal protective equipment needed.
Disinfecting the surface of
the device
A detergent-disinfectant is used to
disinfect the exterior of the machine.
If using a spray or foam detergent-
disinfectant:
o
Spray the product onto a
disposable cleaning pad made of
a non-woven material.
o
Wipe the exterior of the device
and allow to dry.
If using detergent-disinfectant wipes:
o
Wipe the exterior of the device
and allow to dry.
Products which must not be
used
Never use abrasive or highly alkaline
cleaning products, acetone, benzene or
leaded gasoline to clean the device.
Do not use materials such as abrasive
sponges or stiff-bristled brushes.
Never use any pre-disinfecting or
disinfecting
product
containing
aldehydes
and/or
its
derivatives:
formaldehyde, glutaraldehyde, etc.
Do not use any pre-disinfecting or
disinfecting product which is not
effective on Mycobacterium tuberculosis
(BK).
Never disinfect the internal circuit of air
flow using a disinfecting product or with
an Airway Surface Disinfection System
(DSVA).