![Seedburo 9000AG Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/seedburo/9000ag/9000ag_operating-and-service-manual_1232857013.webp)
Seedburo Equipment Company | 9000AG Manual Revision 1.0
13
Section 8
Error Messages:
Error:
Explanation:
Ldn9 0 Loading Zero
The scale will always compare the zero calibration point against the zero on power up. This
message will appear if the scale zero is not stable (could be from wind or vibration) or if there is
an instability problem with the load cell.
0 Ovr Zero Over
You can zero a maximum of 3% of capacity .
Remove all dead load from the scale (pans/seeds etc..) Make sure platter is centered on the pins
of the weighing mechanism. Power on scale.
If it is still in
0 Ovr
than you must enter the calibration menu and re-calibrate the scale.
0 Undr
Zero Under
This message will appear if the initial zero is 100 grams or more
below
the CAL zero point.
Err EEp
Error EEPROM memory is corrupted. Default scale parameters and re-calibrate scale. If
problem persists than you need to replace the main printed circuit board and re-calibrate scale.
Udr Ld
Under Load
Under load message will appear when the scale is 100 grams or less than the CAL zero point.
Our Ld Over Load
Overload message will appear when the scale weight is 103% over initial zero or displayed zero,
whichever comes first.