![Sedecal VET-X APR-VET Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/sedecal/vet-x-apr-vet/vet-x-apr-vet_operation_1232646027.webp)
High Frequency Generators - Veterinary Systems APR-VET
Operation
OM-0118R0
19
AEC RESET
: If the exposure is aborted by the AEC back-up timer, the indicator
on the “
AEC Reset
” push-button blinks accompanied of an audible alarm. Next
exposure is inhibited until the AEC function is reset by pressing the “
AEC Reset
”
push-button. When the Generator is in
“Prep”
mode, the AEC function can not
be reset.
The “
AEC Reset
” push-button may blink when the kVp value, AEC Density and
Film / Screen Combination select a technique that is out of the operative range
with AEC, it inhibits the next exposure. Change any parameter (kVp value, AEC
Density or Film / Screen Combination) in order to obtain a technique enabled
for AEC.
3.3
ANATOMICAL PROGRAMMER (APR)
Anatomical Programmer (APR) selects the values for the radiographic
technique in accordance to the selected Body Region and Thickness in
centimeters. (
Refer to Illustration 3-2)
.
Besides the radiographic parameters, AEC (Density, AEC Fields and
Film/Screen Combination) selections can be assigned to the APR techniques.
APR techniques can be modified and stored again in a non-volatile memory by
the operator.
The APR techniques are intended as a
guide line only.
Accurate exposure
factors depend among other things on the Bucky Grid factors, Table-Top
absorption, Film / Screen Combinations and Film processing.
APR language may be changed just after APR activation by
pressing the “Power ON” push-button. Language selection
remains stored even after the equipment is turned Off. (Only for
Generators equipped with this option).
If an APR technique is to be stored with AEC parameters, a
suitable back-up time (and/or mAs) MUST be stored in the
programme by the operator.
APR DISPLAY
: Shows the different Body Regions (Skull, Abdomen, Pelvis,
Thorax, Extremity, Spine, Exotics or Dental) and the Thickness range in
centimeters of the area to be exposed.
APR MODE SELECTION
: This push-button (C) is used to activate (lighted) or
deactivate the APR mode.
This push-button (A) has not assigned function with this APR mode.
Note
.
Note
.