High Frequency Generators - Veterinary Systems APR-VET
Operation
OM-0118R0
16
3.1.6
RADIOGRAPHIC PARAMETERS
kV DISPLAY
can show:
•
The radiographic kV value selected for the technique.
•
The actual X-ray Tube heat unit value after pressing the “
On
” push-button
(
Refer to Section 3.5
).
•
The error messages during a system fault, preceded by the letter “E”
(a.e., E02) (
Refer to Section 3.7
).
mAs DISPLAY
can show:
•
The radiographic mAs value selected for the technique.
•
When an exposure is made with AEC, it shows the actual mAs at the end
of the exposure whenever the “
Prep
” push-button has not been released.
•
If an exposure is aborted by releasing the Exposure Control (Pedal or
Handswitch) during the exposure, it shows the actual mAs value until the
“
AEC Reset
” push-button is pressed to reset the error condition.
mA DISPLAY
: Shows the radiographic mA value selected for the technique.
Time DISPLAY
can show:
•
The Time value (in seconds) selected for the radiographic technique.
•
When an exposure is made with AEC, it shows the back-up time during
the exposure and the actual Time at the end of the exposure whenever
the “
Prep
” push-button has not been released.
•
If an exposure is aborted by releasing the Exposure Control (Pedal or
Handswitch) during the exposure, it shows the actual Time until the “
AEC
Reset
” push-button is pressed to reset the error condition.
RAD Displays
can also show:
•
The actual Time, the calculated mAs, and the selected kVp and mA
radiographic parameters of the last exposure, with or without AEC, after
pressing the “
AEC Reset
” push-button (values flashing).
•
The exposure counters (
Refer to Section 3.6)
.
kVp
mAs
mA
s