Secure Caregiver CAS-FAM Скачать руководство пользователя страница 3

OPERATION

Affix the alarm monitor and holder securely to the side of the bed 
or chair out of sight and reach of the patient. Simply wrap the 
attachment strap with the rubberized non-slip backing attached 
to the back of the holder tightly around railing or frame of bed or 
wheelchair. 

CAUTION: Do not affix alarm to a PVC rail. The 

strap may permanently stain the rail. Holder placement 

should be easily accessible to staff. Secure excess pad cord 

to avoid damaging the cord.
Mat Mode

When Mat Mode is selected, the alarm monitor will trigger when the 
patient steps on the floor mat sensor, alerting caregiver of unsafe 
egress or wandering. Floor mat sensors are typically used to prevent 
patient wandering. Place floor mat sensor next to the patient bed, in 
a doorway, or any other area that wandering is unwanted. 

Pad Mode

When patient initially lies on the bed sensor pad or sits on the chair 
sensor pad, the alarm monitor will generate a short beep tone to 
confirm that the sensor pad is active. The alarm monitor will trigger 
when the patient attempts to get out of the bed, chair, or wheelchair, 
alerting caregiver of unsafe egress or wandering. The alarm monitor 
alert sound will trigger immediately after patient weight is removed 
from the sensor pad. 
When caregiver removes patient from sensor pad for whatever 
reason, the alarm monitor alert sound will trigger immediately. 

To stop alarm from sounding press and release the Test/

Reset button 3 times within 2 seconds. 

The alarm monitor 

will automatically reset, silencing the alert sound, when the patient 
returns weight to the sensor pad. 

THE ALARM MONITOR CAN BE RESET (SILENCED) BY 

PRESSING THE RED RESET BUTTON (located on the front 

of the alarm monitor) THREE TIMES IN SEQUENCE. 

The three 

push reset feature prevents patient from tampering with the alarm 

monitor. Only the caregiver(s) should know about the three push 

reset feature to prevent patient tampering. 

NOTE: If using the Secure

®

 Bed Pads, Secure

®

 Chair Pads or 

Secure

®

 Seat Belts see instructions printed on pad or included 

with seat belt. 

CAUTION! ALWAYS remove sensor pad pin 

from alarm by pulling straight up on the white “PULL” tab. 

DO NOT attempt to remove pin any other way or you may 

damage pin and void warranty.
Three Second Delay Mode (+3S)

Functions the same as Pad Mode with an added three second alert 
sound delay when bed/chair sensor pad is triggered. May help 
prevent false alarms when monitoring patients who tend to move 

around on bed or chair. 

Nurse call

Plug the Universal Nurse Call Button (included) into the ¼” input 
located on the bottom side of the Fall Alarm Monitor. Pressing the 
nurse call button once will send a signal directly to the Wireless 
Caregiver Pager (Model No. CAS-PGR). The fall alarm monitor unit 
will generate confirmation beep tone to indicate the nurse call button 
has been pressed by the patient. The nurse call button should be 
always be within reach of the patient. 
NOTE: Universal Nurse Call Button can only be used only when the 
Wireless Caregiver Pager (Model No. CAS-PGR) is paired with the 
Fall Alarm Monitor.

Test & Reset

Test: Pressing the Reset/Test button once while the Fall Alarm 
Monitor is in standby mode will produce a short beep tone and 
a wireless signal will be sent to the Wireless Caregiver Pager to 
confirm that it is properly paired with the Fall Alarm Monitor. Note 
that the Wireless Signal Transmission On/Off Switch must be in 
the “On” position for the Reset/Test button to activate the Wireless 
Caregiver Pager during testing.
Reset: To stop alarm from sounding press and release the Test/
Reset button 3 times within 2 seconds.
The alarm monitor and sensor pad should be checked prior to each 
use for proper functioning. Please note: The system can be checked 
while resident is on the pad by simply pressing “Test/Reset” button. 
The alarm will beep twice. 

DO NOT ASSUME! TEST EACH TIME 

THE SYSTEM IS USED.

12 MONTH LIMITED WARRANTY

Personal Safety Corporation warrants this product to be free from 
factory defects in materials and workmanship for a period of 12 
months from the date of purchase. Void if this product has been 
misused, abused or punctured. No warranty of responsibility of use 
expressed or implied.

CAUTION!

This product is not a restraint. It will not prevent a resident from 
getting out of their bed or chair.
This product is designed as a monitor, alerting the staff when the 
resident has risen from their bed or chair
This product is to be used in conjunction with a total fall prevention 
program.

 Revised: 07/26/17

FALL ALARM  

MONITOR &  

PAGER ALERT 

Continued

Please see our complete line of Secure

®

 fall/wandering prevention & patient safety products  

at www.SecureSafetySolutions.com 

Highest Quality...Lowest Prices...Guaranteed!