1.
Power
• Switch to on position to turn on alarm monitor.
• Switch to off position when alarm monitor is not in use to conserve batteries and avoid false
alarm noise.
2.
Transmitter On/Off switch
• Switch to on position to turn on wireless mode. When wireless mode is on, a wireless signal is sent
to the optional Wireless Caregiver Pager (Model No. CAS-PGR) when alarm monitor is triggered to alert
caregiver remotely
• Switch to off position to turn off wireless signal function
and conserve battery life.
3.
Alert Sound selector
• Select desired alert sound by sliding the switch
to • or •• position.
4.
Alert Light On/Off
• Turn the flashing alert light on or off .
5.
Mat Mode or Pad Mode Selection
(Mat, Pad, +3S)
•
“Mat” Mode (floor mat sensor):
Select “Mat” mode
when using a floor mat sensor next to a bed or in a
doorway. The alarm monitor will trigger when the patient
steps on the mat, alerting caregiver of unsafe egress or
wandering.
6.
Alert Sound Volume control
• Select High
II
/ Low
I
/ Mute to choose the desired alert
sound volume level.
WARNING! Mute should only be selected when
the Fall Alarm Monitor is paired with the Wireless
Caregiver Pager.
7.
New RF Code
• Press the
button using a pen or other pointed object
to generate a new RF code and eliminate any interference
caused by another alarm monitor or equipment being used
in close proximity.
EXIT
Personal Safety Corporation • Hiawatha, Iowa 52233 • 800.373.2873 • www.SecureSafetySolu
tions.com
SETTINGS
“Pad” Mode (bed or chair sensor Pad):
Select “Pad” mode
when using a bed or chair sensor pad. The alarm monitor will
trigger when patient weight is removed from the pad, alerting
caregiver of unsafe egress.
“Pad + 3S” Mode:
Adds three second alert delay from the
time patient weight is removed from bed or chair sensor pad).
May be helpful in reducing false alarms with patients who tend
to move around in the bed or chair.
Pairing Alarm Monitor (Model No. CAS-FAM)
with optional Wireless Caregiver Pager
(Model No. CAS-PGR)
• The CAS-FAM Alarm Monitor works as a standalone fall alarm
monitor or it can be paired with the CAS-PGR Wireless Caregiver
Pager and/or CAS-WCR Wireless Caregiver Receiver. To pair with
the pager/receiver, please follow the instructions included with the
pager/receiver to activate pairing mode. When the pager/receiver
is in pairing mode, simply activate the alarm monitor by pressing
down on the connected sensor pad or floor mat to pair with the
pager/receiver. Alternatively, the connected Universal Nurse Call
Button may be pressed to pair the alarm monitor with the pager/
receiver.
FALL ALARM
MONITOR &
PAGER ALERT
Continued