4. Maintenance
Only qualified technicians are allowed to perf
rr o
ff rm maintenance on the UrSpring.
Non-expert and insuff
ff icient maintenance will fo
ff rf
rr eit any warranty rights.
4.1 Maintenance schedule
Part
no.
Description
Qty
Serv
rr ice
interv
rr al
Task
Task description
10963
Pressure relief
automatics
1
36
month
Calibrate
Sensors
Re-calibrate sensors through the
user interf
rr ace.
11351
Flow sensor
1
36
month
Function
test
Open tap downstream and
check the flow displayed on your
unit. If those matches, test is
completed.
10787
Flush-Valve
1
36
month
Function
test
Check function by perf
rr o
ff rming a
manual flush through the control
of the UrSpring.
10646
Quick-Connect-
Adapter
3
36
month
Check fo
ff r
leaks
10788
Only UrSpring
HotWater Filter
1
12
month
Replace
10545
all otherr
UrSpring Filters
1
36
month
Replace
Page 32/50