Preparatory work in the office
32
8.7
Registering devices in the SebaDataView-3 software
8.7.1
Introduction
All GT-3-1 and all loggers that are to be used for the upcoming measurement must have
been registered in the SebaDataView-3 software.
You can register the devices in the office, or do this directly on site.
Observe the following points:
•
All devices in the network (i.e. all GT-3-1 used and all loggers) must be combined to
form a shared zone.
•
A specific group must be created for every GT-3-1 that is used for the
measurement. The loggers that are to be connected to the GT-3-1 concerned are
added to this group.
•
A maximum of 3 loggers can be added to a GT-3-1 group.
Example:
8.7.2
Selecting a zone or creating a new one
Select the zone to which the GT-3-1 and the loggers concerned should be added in the
directory tree, or create a new zone.
To create a new zone, proceed as follows:
Step Description
1
Mark the folder in which the new zone should be created.
2
Open the
Directory
tab in the menu bar.
3
Click
New
in the
Zone
segment.
4
In the window which opens, give the new zone a name and add a comment.
Confirm the inputs using
OK
.
Result:
The new zone has now been created in the database and will appear in
the directory tree.
Important information
Shared zone
for the GT-3-1 network
Group “Group I”,
assigned GT-3-1 “9964” and the loggers
that are to be connected to it
Group “Group II”,
assigned GT-3-1 “9967” and the loggers
that are to be connected to it
Содержание KMT Sebalog GT-3-1
Страница 2: ...Consultation with SebaKMT 2 ...
Страница 20: ...Data security 20 ...
Страница 22: ...First use 22 ...
Страница 26: ...Inserting the SIM card changing the battery 26 ...
Страница 28: ...Switching the GT 3 1 on off 28 ...