Section
2
Installation of Analyzer
2.1
Unpacking
Unpack the carton case and check with Packing List, and prepare to install the analyzer if
everything is right. If you have any questions, please contact with our after-sales service team.
Meanwhile, please write the guarantee and send it back, which is convenient for us to follow
and track your service. (If send email, please email to: [email protected])
Instrument work environment : Temperature: 10-32
℃
,
Relative Humidity ≤85%
Note
Do not place the device in following environment :
Humidity, corrosive gas, dusty, strong electromagnetic interference;
1)
Crowd and non-aeration;
2)
Non-steady worktable;
3)
Direct sunlight.
4)
Requirements for power supply:
AC100-240V, 50-60Hz,
, 105W Max
The analyzer can not use the same power supply socket with some high-power
equipment including centrifugal, refrigerator, and oven, etc.
The power supply should be connected with three-cores wire and and good grounding ;
the voltage between the neutral line and earth line
<
5V; or choose two core power cord
with power adapter provided by Seamaty.
2.2 Installation Procedure
2.2.1 Set up analyzer
1)
Remove the analyzer from the shipping carton case and place in a stable worktable
Check if the exterior case of analyzer is intact
2)
3)
Plug the power jack into the analyzer, then plug the detachable power supply cord
into the power adapter and into a grounded electrical outlet.
Turn on the switch on the back of analyzer, indicator light on, the display will show
4)
the home screen.
Please keep safe of accessories have not been used.
Instructions