3. specification
4. operation
Warning!
ensure that you read, understand and apply the safety and operational instructions in Section.1 before connecting the
analyser. only when you are sure that you understand the procedures, is it safe to proceed with testing.
Warning!
Risk of electrocution. High voltage circuits, both AC and dC are very dangerous and should be measured with great care.
Remember to turn meter off when measurement is completed.
note:
If ‘ol’ (over limit) appears in the display during a measurement, the value exceeds the range that is selected. Change to a higher range.
note:
on some low AC and dC ranges, with the test leads not connected to a device, the reading may show a random fluctuating reading.
This is normal and is caused by the high input sensitivity. The reading will stabilise and give an accurate measurement when
connected to a circuit.
Dc voltage
(Auto Ranging)
range resolution accuracy
400.0mV 0.1mV
± 0.5% of reading ± 3 digits
4.000V 1mV
40.00V
10mV
±1.5% of reading ± 2 digits
400.0V 100mV
600V
1V
± 1.8% of reading ± 2 digits
Input Impedance: 10M
Ω.
Maximum Input: 600Vdc.
range resolution accuracy
400.0mV
0.1mV
± 1.5% of reading ± 5 digits
4.000V
1mV
± 1.0% of reading ± 3 digits
40.00V
10mV
± 1.5% of reading ± 3 digits
400.0V 100mV
600V
1V
± 2.0% of reading ± 4 digits
ac voltage
(Auto Ranging)
Input Impedance: 10M
Ω.
Frequency Range: 50 to 60Hz.
Maximum Input: 600Vac rms.
range resolution accuracy
0.5%
~
99.0%
0.1%
± 2.0% of reading ± 5 digits
Duty cycle
Pulse Width: >100us, <100ms
Sensitivity: > 5V rms
Frequency width: 5Hz - 100kHz
cylinder range
resolution accuracy
4CYl 0~90.0°
5CYL
0~72.0° 0.1°
± 2.0% of reading ± 4 digits
6CYl 0~60.0°
8CYl 0~45.0°
Dwell angle
range resolution accuracy
4.000kHz 1Hz
40.00kHz
10Hz
± 1.5% of reading ± 3 digits
400.0kHz 100Hz
4.000MHz 1000Hz
40.00MHz
1kHz
± 2.0% of reading ± 4 digits
frequency
(Auto Ranging)
Sensitivity: >5V RMS MIn up to 4.0000MHz range.
>15V RMS MIn on 40.00MHz range.
range resolution accuracy
4nF
1pF
± 5.0% of reading ± 7 digits
40nF
10pF
± 5.0% of reading ± 7 digits
400nF 0.1nF
4uF
1nF
± 3.0% of reading ± 5 digits
40uF 10nF
400uF 0.1uF
4mF
0.001mF
±10.0% of reading ±10 digits
40mF 00.01mF
capacitance
(Auto Ranging)
range resolution accuracy
400.0Ω
0.1Ω
± 1.2% of reading ± 4 digits
4.000kΩ
1Ω
± 1.0% of reading ± 2 digits
40.00kΩ
10Ω
± 1.2% of reading ± 2 digits
400.0kΩ
100Ω
4.000MΩ
1kΩ
40.00MΩ
10kΩ
± 2.0% of reading ± 3 digits
resistance
(Auto Ranging)
open Circuit Voltage: 3.0V dc typical.
range resolution accuracy
1.0mA typical 1mV
± 5% of reading ± 15 digits
Diode test
range resolution accuracy
-30°C~1000°C 1°C
± 3.0% of reading ± 5°C / 8°F
-22°F~1832°F 1°F
(Meter only, probe accuracy
not included).
temperature
Sensor: Type K Thermocouple
audible continuity
Audible threshold:
less than 35
Ω
Test Current:
< 1mA dc typical
range resolution accuracy
40.00mA 10uA
400.0mA
100uA
± 1.8% of reading ± 5 digits
20A
10mA
± 3.0% of reading ± 7 digits
ac current
overload Protection: 0.5A/250V and 20A/250V Fuse.
Frequency Range: 50 to 60Hz
Maximum Input: 40mA ac rms or 400mA dc on uA/mA ranges,
20A ac rms or dc on 20A range.
range resolution accuracy
1.0~20.0ms
0.1ms
± 2% of reading ± 20 digits
pulse Width
range
resolution accuracy
RPM 4 600~4000RPM
1RPM
1000~12000RPM 10RPM
(x 10RPM)
±2% of rdg ± 4 digits
RPM 2 300~4000 RPM
1RPM
& dIS 1000~6000RPM 10RPM
(x 10RPM)
effect Reading: >600RPM
rpm
(Tach)
accuracy is given at 18°c to 28°c (65°f to 83°f) with
less than 70% relative humidity.
range resolution accuracy
40.00mA 10uA
400.0mA
100uA
± 1.5% of reading ± 3 digits
20A
10mA
± 2.5% of reading ± 5 digits
Dc current
overload Protection: 0.5A/250V and 20A/250V Fuse.
Maximum Input:400mA ac rms or 400mA dc on uA/mA ranges,
20A ac rms or dc on 20A range.
TA303 Issue 2 (HF) 03/10/18
Original Language Version
© Jack Sealey limited
function
reD terminal
input limit
dC/AC Volts, ohms, Continuity, diode, CAP.,
Type-K TEMP., Hz, % Duty,
Ms dwell, RPM
V
ΩHz% etc.
600Volts AC dC
Current AC/dC 400mA
400mA
400mA dC/AC
Current AC/dC 20A
20A
*20A dC/AC
* 20 Amp measurement for 30 seconds maximum.
1
; ohms cannot be measured if voltage is present, ohms can be measured only in a non-powered circuit. However, the analyser is protected to 600 volts.